HPLC Analysis of Benzodiazepines on Ascentis® Express Phenyl-Hexyl, Methanol Mobile Phase

材料
Standard
製品番号
詳細
価格
分析カラム
製品番号
詳細
価格
CONDITIONS
column
Ascentis Express Phenyl-Hexyl, 5 cm x 4.6 mm I.D., 2.7 μm particles (53348-U)
mobile phase
[A] water; [B] methanol; (35:65, A:B)
flow rate
1.4 mL/min
pressure
3600 psi (248 bar)
column temp.
35 °C
detector
UV, 250 nm
injection
2 μL
sample
0.1 g/L in 20% methanol
詳細
アナリシスノート
The Phenyl-Hexyl phase of the Ascentis Express product line displays the best selectivity of this sample; selectivity is markedly affected by choice of organic modifier. The application also demonstrates feasibility of fast LC with Ascentis Express by virtue of low backpressure and high efficiency.