Passa al contenuto
Merck

HPLC Analysis of Paraquat and Diquat on SUPELCOSIL LC-18 after SPE using ENVI-8 DSK

HPLC Analysis of Paraquat and Diquat on SUPELCOSIL™ LC-18 after SPE using ENVI™-8 DSK application for SPE, application for HPLC

Materiali

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Colonna per HPLC SUPELCOSIL LC-18

5 μm particle size, L × I.D. 15 cm × 4.6 mm

Prodotti correlati

N° Catalogo
Descrizione
Determinazione del prezzo

ENVI-8 DSK SPE Disk

diam. 47 mm, pk of 24

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

drinking water adjusted to pH to 10.

sample preparation

ENVI-8 DSK, 47 mm (57172)

SPE tube/cartridge

ENVI-8 DSK SPE Disk, 47 mm (57172)

condition

10 mL acetonitrile; 2 × 10 mL reagent water; 10 mL conditioning solvent A (5 g cetyl trimethyl ammonium bromide and 5 mL conc. ammonium hydroxide in 500 mL DI water, dilute to 1 L); (2 × 10 mL reagent water; 10 mL conditioning solvent B (10 g hexanesulfonic acid, sodium salt and 10 mL conc. ammonium hydroxide in 250 mL DI water, dilute to 500 mL))

sample addition

250 mL water 100 mL/min

elution

0.5-1.0 mL acetonitrile (to cover/solvate disk); 2 × 4 mL eluting solution (13.5 mL orthophosphoric acid and 10.3 mL diethylamine in 500 mL DI water, dilute to 1 L)

column

SUPELCOSIL LC-18, 15 cm × 4.6 mm I.D., 5 μm particles (58230-U)

mobile phase

3.5 mL triethylamine and 1.0 g 1-hexane-sulfonic acid, sodium salt to 800 mL DI H2O add orthophosphoric acid to pH 2.5, dilute to 1 L

flow rate

1.0 mL/min

column temp.

35 °C

detector

photodiode array, diquat - 308 nm, paraquat - 257 nm

injection

100 μL

Descrizione

Applicazioni

Refer to US EPA Method 549.1 for full details

Sample Pre-treatment:
250 mL drinking water. Adjust water sample to pH to 10.5 ± 0.2 with 10% sodium hydroxide or 10% hydrochloric acid

Note legali

ENVI is a trademark of Sigma-Aldrich Co. LLC
SUPELCOSIL is a trademark of Sigma-Aldrich Co. LLC