Passa al contenuto
Merck

GC Analysis of Peppermint Essential Oil on SLB®-IL59

GC Analysis of Peppermint Essential Oil on SLB®-IL59 suitable for GC

Materiali

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

SLB®-IL59 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

SLB-IL59, 30 m x 0.25 mm I.D., 0.20 μm (28891-U)

oven

50 °C, 3 °C/min to 300 °C

inj. temp.

220 °C

detector

FID, 300 °C

carrier gas

helium, 30 cm/sec

injection

1 μL, 100:1 split

sample

Peppermint oil purchased from Misitano&Stracuzzi S.p.a. (Messina, Italy) diluted 1:10 in n-hexane

Descrizione

Note legali

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany