G005284
GC Analysis of Glycols in Water on SPB®-1000 after SPE using ENVI-Carb™ Plus
suitable for GC, application for SPE
About This Item
Prodotti consigliati
tecniche
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: tap water spiked at 2 μg/mL with glycols
SPE tube/cartridge: ENVI-Carb Plus reversible cartridge, 400 mg/ 1mL (54812-U)
condition: 1 mL methylene chloride, 2 x 2 mL methanol, 3 mL deionized water (cartridge not allowed to dry during conditioning)
sample addition: 5 mL spiked tap water
drying: 10 min, 10 mm mercury
elution: 2 mL 80:20 methanol:methylene chloride
column: SPB-1000, 30 m x 0.25 mm I.D., 0.25 μm (24313)
oven: 50 °C (1 min), 8 °C/min to 200 °C (7 min)
inj. temp.: 220 °C
detector: FID, 220 °C
carrier gas: helium, 1.5 mL/min. constant
injection: 1 μL, splitless (splitter open at 0.75 min.)
liner: 4 mm I.D. FocusLiner™ inlet liner with taper
Compatibilità
application for SPE
applicazioni
environmental
Note legali
Analita
Colonna analitica
Colonnina o piastra per SPE
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.