Passa al contenuto
Merck
Tutte le immagini(1)

Documenti fondamentali

Z763780

BeadBug prefilled tubes, 2.0 mL capacity

with 1.0 mm Zirconium beads, triple-pure, high impact

Sinonimo/i:

homogenization microtubes

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali

Scegli un formato

50 EA
194,00 €

194,00 €


Per informazioni sulla disponibilità, contatta il Servizio Clienti.


Scegli un formato

Cambia visualizzazione
50 EA
194,00 €

About This Item

Codice UNSPSC:
41106300
NACRES:
NB.22

194,00 €


Per informazioni sulla disponibilità, contatta il Servizio Clienti.

Sterilità

non-sterile

Confezionamento

pack of 50 ea

Produttore/marchio commerciale

Benchmark Scientific D1032-10

tecniche

PCR: suitable

Cerchi prodotti simili? Visita Guida al confronto tra prodotti

Descrizione generale

These prefilled tubes fit the BeadBug microtube homogenizer, and come with standard acid-washed glass beads, triple-pure zirconium beads, or stainless steel beads.
The triple-pure beads are molecular biology grade, and are acid washed, heat treated and tested free of nuclease.
The stainless steel beads are acid-washed as well.
All tubes are non-skirted with screw-caps

Note legali

BeadBug is a trademark of Benchmark Scientific, Inc.

Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documenti section.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Domande

1–2 di 2 domande  
  1. Can these beads be reused after washing and sterilization?

    1 risposta
    1. The beads are reusable, but the tubes should not be reused. The directions for washing the zirconium beads are as followed:

      The glass and zirconium beads will eventually break down, so inspect before use for cracks and/or broken pieces. (The speed at which this happens will depend on the hardness of the sample and length of homogenization.) Collect the beads in water and rinse them multiple times with deionized water, drain, and autoclave. The beads should then be baked in an oven at 500ºF for an hour (or 300ºF for 2 hours, or 180ºF for several hours) to render them protease and nuclease-free. If after baking, the beads are not free-flowing (they are sticking together), then they are not clean and the process should be repeated.

      Utile?

  2. Is product Z763780, BeadBug prefilled tubes (Benchmark Scientific product D1032-10) autoclaveable? If so do the caps need to be removed first?

    1 risposta
    1. The tubes can be autoclaved. It is recommended to loosen the caps if autoclaving. However, this product is certified to be nuclease and protease free and is filled in a sterile process that is tested to be free of stray DNA. Therefore, there is no reason to autoclave them. In fact, autoclaving may introduce contaminants.

      Utile?

Recensioni

Nessuna valutazione

Filtri attivi

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.