KHGEG003FB3
Capsule Opticap® XL, membrana Millipore Express® SHC
Opticap® XL 3, inlet connection diam. 3/4 in., pore size 0.5/0.2 μm, cartridge nominal length 4.7 in. (11.9 cm)
Sinonimo/i:
Opticap Gamma Compatible XL 300 Millipore Express SHC 0.5/0.2 μm 3/4 in. TC-9/16 in. HB with Bell
About This Item
Prodotti consigliati
Materiali
polyethersulfone
polyethylene support
polypropylene (gamma-stable)
polypropylene housing
polypropylene vent cap (Gamma stable)
polysulfone
silicone seal
Livello qualitativo
Conformità normativa
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
Descrizione
19 mm (3/4 in.) Sanitary Flange Inlet and 14 mm (9/16 in.) Hose Barb Outlet with filling bell
Sterilità
non-sterile
Compatibilità con la sterilizzazione
gamma compatible
Nome Commerciale
Opticap® XL 3
Caratteristiche
hydrophilic
Produttore/marchio commerciale
Millipore Express®
Parametri
≤1.9 mL/min air diffusion at 2.8 bar (40 psig) and 23 °C (in water)
1.0 bar max. differential pressure (15 psid) at 80 °C (Forward)
1.0 bar max. inlet pressure (15 psi) at 80 °C
100 psi max. differential pressure (6.9 bar) (intermittent)
2.1 bar max. differential pressure (30 psid) at 25 °C (Reverse; intermittent)
2.75 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)
5.5 bar max. inlet pressure (80 psi) at 23 °C
6.9 bar max. differential pressure (100 psid) at 25 °C (Forward; intermittent)
6.9 bar max. inlet pressure (100 psi) at 23 °C (intermittent)
80 psi max. differential pressure (5.5 bar) (continuous)
80 psig max. inlet pressure
tecniche
sterile filtration: suitable
Lunghezza
11.9 cm (4.7 in.)
Larghezza
2.2 in.
Lunghezza nominale della cartuccia
4.7 in. (11.9 cm)
Diametro
5.6 cm (2.2 in.)
Superficie filtrante
290 cm2
Diam. connessione ingresso
3/4 in.
Amp. ingresso/ uscita
11.9 cm (4.7 in.)
Diam. connessione uscita
9/16 in.
Impurezze
≤0.25 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
Matrice
Millipore Express® SHC
pore size
0.5/0.2 μm pore size
Punto di bolla
≥4000 mbar (58 psig), air with water at 23 °C
Raccordi
1/4 in. drain/vent hose barb (with double O-ring seal)
inlet sanitary flange
outlet hose barb (with Bell)
Descrizione generale
Confezionamento
Altre note
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer installation section of OPTICAP® XL Capsules and OPTICAP® XLT Capsules user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Nota sulla preparazione
Gamma compatible to 45 kGy; 3 autoclave cycles of 60 min @ 123 °C; not in-line steam sterilizable
Risultati analitici
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
After sterilization and a controlled water flush of 2.0 L, samples exhibited less than 500 ppb TOC per USP <64> and less than 1.3 µS/cm per USP <64> at 25 °C.
Note legali
Esclusione di responsabilità
Non trovi il prodotto giusto?
Prova il nostro Motore di ricerca dei prodotti.
Certificati d'analisi (COA)
Cerca il Certificati d'analisi (COA) digitando il numero di lotto/batch corrispondente. I numeri di lotto o di batch sono stampati sull'etichetta dei prodotti dopo la parola ‘Lotto’ o ‘Batch’.
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.