G005540
HPLC Analysis of Dermorphin in Horse Plasma on Ascentis® Express F5 after SPE with HybridSPE®
application for HPLC, application for SPE
About This Item
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: horse plasma spiked at 10 μg/mL
SPE tube/cartridge: [TM"HybridSPE"]-Phospholipid Plate, 50mg/96-Well (575656-U)
sample addition: 3:1, (1% formic acid in acetonitrile); plasma
column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: [A] acetonitrile [B] 20 mM ammonium formate; (90:10, A:B); pH 4.5 with formic acid, after mixing A & B
flow rate: 0.4 mL/min
pressure: 1305 psi (90 bar)
column temp.: 35 °C
detector: MS, ESI(+), TIC, m/z 100-1000
injection: 1 μL
suitability
application for HPLC
application for SPE
application(s)
forensics and toxicology
Analysis Note
Legal Information
analytical column
mobile phase component
Vyberte jednu z posledních verzí:
Osvědčení o analýze (COA)
Je nám líto, ale pro tento produkt momentálně nemáme COA k dispozici online.
Potřebujete-li pomoc, obraťte se na Zákaznická podpora
Již tento produkt vlastníte?
Dokumenty související s produkty, které jste v minulosti zakoupili, byly za účelem usnadnění shromážděny ve vaší Knihovně dokumentů.
Náš tým vědeckých pracovníků má zkušenosti ve všech oblastech výzkumu, včetně přírodních věd, materiálových věd, chemické syntézy, chromatografie, analytiky a mnoha dalších..
Obraťte se na technický servis.