Přejít k obsahu
Merck
Všechny fotografie(1)

Hlavní dokumenty

G005391

HPLC Analysis of Diet Soda Spiked with Neotame on Ascentis® Express RP-Amide

application for HPLC

Přihlásitk zobrazení cen stanovených pro organizaci a smluvních cen


About This Item

technique(s)

HPLC: suitable

test parameters

column: Ascentis Express RP-Amide, 3 cm x 4.6 mm I.D., 2.7 μm particles (53921-U)
mobile phase: [A] 100 mM ammonium acetate, pH 5.6 with acetic acid; [B] water; [C] acetonitrile
gradient: 20% A constant; 75 to 20% B, 5 to 60% C in 1 min; held at 20% B, 60% C for 0.1 min
flow rate: 3 mL/min
pressure: 3336 psi (230 bar)
column temp.: 40 °C
detector: UV, 214 nm
injection: 1 mL/min
sample: diet soda spiked with neotame 350 μg/mL

suitability

application for HPLC

application(s)

food and beverages

Analysis Note

This application demonstrates the suitability of Ascentis Express RP-Amide for the efficient separation of common drink additives used in diet cola.

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

analytical column

standard

Č. produktu
Popis
Stanovení ceny

Vyberte jednu z posledních verzí:

Osvědčení o analýze (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Dokumenty section.

Potřebujete-li pomoc, obraťte se na Zákaznická podpora

Již tento produkt vlastníte?

Dokumenty související s produkty, které jste v minulosti zakoupili, byly za účelem usnadnění shromážděny ve vaší Knihovně dokumentů.

Navštívit knihovnu dokumentů

Náš tým vědeckých pracovníků má zkušenosti ve všech oblastech výzkumu, včetně přírodních věd, materiálových věd, chemické syntézy, chromatografie, analytiky a mnoha dalších..

Obraťte se na technický servis.