G004432
HPLC Analysis of Statins in Rat Plasma on Ascentis® Express C18 after SPE using Supel™-Select HLB
application for SPE, application for LC-MS
About This Item
técnica(s)
LC/MS: suitable
solid phase extraction (SPE): suitable
parâmetros de teste
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rat plasma spiked with provastin and atorvastin at 5 and 100 ng/mL
SPE tube/cartridge: Supel-Select HLB SPE, 30 mg/1 mL (54181-U)
condition: 0.5 mL methanol:acetonitrile (1:1, v/v); 0.5 mL DI water
sample addition: 0.5 mL
washing: 0.5 mL 5% methanol
elution: 0.5 mL methanol:acetonitrile (1:1, v/v)
column: Ascentis Express C18, 5 cm x 2.1 mm, 2.7 μm particles (53822-U)
eluate post-treatment: evaporate to dryness with nitrogen at 37 °C, reconstitute in 0.5 mL DI water
mobile phase: 0.1% acetic acid diluted in 60% methanol
flow rate: 0.2 mL/min
column temp.: 35 °C
detector: ABI 3200 QT; ESI(+), MRM (423.3/321.3, 423.3/101.0, 557.3/397.2 and 557.3/453.4)
injection: 5 μL
adequação
application for LC-MS
application for SPE
aplicação(ões)
clinical
Informações legais
coluna analítica
tubo ou placa SPE
Choose from one of the most recent versions:
Certificados de análise (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica