96-0327
GC Analysis of Solvents in Water on SPB®-1 SULFUR after SPME using 75 μm Carboxen/PDMS Fiber
suitable for GC, application for SPME
About This Item
técnica(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
parâmetros de teste
sample/matrix: solvents at 20 ppb in 4 mL water + 25% NaCl in 4 mL vial
SPME fiber: carboxen/PDMS, 75 μm (57318)
extraction: immersion, 10 min, rapid stirring
desorption process: 270 °C, 5 min
column: SPB-1 SULFUR, 30 m × 0.32 mm I.D., 4.0 μm film (24158)
oven: 50°C (2 min) to 150°C at 10°C/min
inj. temp.: splitless (closed 2 min), 260°C, 0.75 mm I.D. liner
detector: FID
carrier gas: helium, 30 cm/sec
adequação
application for SPME
aplicação(ões)
chemicals and industrial polymers
environmental
forensics and toxicology
Informações legais
Analito
Fibra para SPME
coluna analítica
Choose from one of the most recent versions:
Certificados de análise (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica