80-207
GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Water on a 5% Phenyl Column after SPE using Supelclean™ ENVI-18
suitable for GC, application for SPE
About This Item
técnica(s)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
parâmetros de teste
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: water spiked with PAHs
SPE tube/cartridge: Supelclean ENVI-18, 500 mg/6 mL (57064)
condition: 2 x 6 mL toluene:methanol (10:1); 6 mL methanol; 6 mL DI water
sample addition: 250 mL water sample at 10 mL/min
drying: 10 min under vacuum
elution: 2 x 1 mL toluene:methanol (10:1)
column: 5% phenyl polysiloxane, 30m × 0.25 mm I.D., 0.25 μm film (Supelco equivalent, SLB-5ms, 28471-U)
oven: 70 °C (2 min) to 280 °C at 8 °C/min
detector: FID, 310 °C
carrier gas: helium
injection: 1 μL
adequação
application for SPE
aplicação(ões)
environmental
food and beverages
Informações legais
Analito
tubo ou placa SPE
Escolha uma das versões mais recentes:
Certificados de análise (COA)
Lamentamos, não temos COA para este produto disponíveis online no momento.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica