796-0685
GC Analysis of Volatiles in Water on Equity-1 after SPME using 30 μm PDMS Fiber
suitable for GC, application for SPME
About This Item
técnica(s)
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
parâmetros de teste
sample/matrix: 0.7 mL water spiked with 2-600 μg/liter each analyte, 0.25 g NaCl added
SPME fiber: 30 μm PDMS (57309)
extraction: headspace, 12 sec (no stirring), ambient temp.
desorption process: 2-3 min, 250 °C (0.75-1.0 mm I.D. inj. port liner)
column: Equity-1, 10 m × 0.20 mm I.D., 1.2 μm (available on request)
oven: 70 °C (0.2 min) to 180 °C at 50 °C/min
detector: FID, 250 °C
adequação
application for SPME
aplicação(ões)
environmental
food and beverages
Informações legais
Analito
Fibra para SPME