Pular para o conteúdo
Merck
Todas as fotos(1)

Documentos Principais

SN03A47FF3

Millipore

Optiscale® Filtro de cápsula Polygard® CN Durapore

filtration area 17.7 cm2, pore size 0.3 μm

Sinônimo(s):

Optiscale Capsule Polygard-CN 0.3 μm 3/4 in. TC/TC

Faça loginpara ver os preços organizacionais e de contrato


About This Item

Código UNSPSC:
23151806
eCl@ss:
32031610
Preço e disponibilidade não estão disponíveis no momento.

Materiais

PVDF vent cap
fluoroelastomer FKM seal
polycarbonate
polypropylene
polypropylene support

Nível de qualidade

conformidade reg.

meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)

fabricante/nome comercial

OptiScale®

Parâmetros

25 °C max. inlet temp.
5.5 bar max. inlet pressure (80 psi) at 25 °C
80 psig max. inlet pressure

Diâmetro

6.9 cm (2.7 in.)

área de filtração

17.7 cm2

diâm. conexão de entrada

3/4 in.

diam. conexão de saída

3/4 in.

Impurezas

≤0.25 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)

Matriz

Polygard®-CN

tamanho de poro

0.3 μm nominal pore size
0.3 μm pore size

entrada

sample type liquid

conector

inlet sanitary flange
19 mm (3/4 in.) inlet/outlet sanitary flange
outlet sanitary flange

Procurando produtos similares? Visita Guia de comparação de produtos

Categorias relacionadas

Descrição geral

Device Configuration: Capsule

Características e benefícios

Format: Single Layer

Outras notas

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
Directions for Use:
  • Organism Retention: Bacterial
  • Mode of Action: Filtration (size exclusion)
  • Appliication: Protein Purification
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer to the installation and wetting sections of OptiScale® 47 Capsulle Filter with Polygard® CN Media user guide
  • Storage Statement: Store in original packaging at room temperature
  • Disposal Statement: Dispose of in accordance with applicable federal, state, and local regulations

Nota de preparo

Sterilization Method
3 autoclave cycles of 60 min @ 126 °C; not in-line steam sterilizable

Nota de análise

Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥100 mL

Informações legais

OptiScale is a registered trademark of Merck KGaA, Darmstadt, Germany
POLYGARD is a registered trademark of Merck KGaA, Darmstadt, Germany

Certificados de análise (COA)

Busque Certificados de análise (COA) digitando o Número do Lote do produto. Os números de lote e remessa podem ser encontrados no rótulo de um produto após a palavra “Lot” ou “Batch”.

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Questions

Reviews

No rating value

Active Filters

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica