Cápsula Millipore Express® SHF, Opticap® XLT
Opticap® XLT 30, cartridge nominal length 30 in. (75 cm), filter diam. 19.8 cm (7.8 in.)
About This Item
Materiais
polyester support
polypropylene device
polypropylene housing
polypropylene vent cap
polysulfone device
silicone seal
Nível de qualidade
conformidade reg.
meets FDA 21 CFR 177-182 (all component materials)
esterilidade
non-sterile
Compatibilidade de esterilização
gamma compatible
linha de produto
Opticap® XLT 30
Características
hydrophilic
fabricante/nome comercial
Millipore Express®
Parâmetros
1.0 bar max. inlet pressure (15 psi) at 80 °C
1000 mbar max. differential pressure (15 psi) at 80 °C (Forward)
2.75 bar max. inlet pressure (40 psi) at 60 °C
2100 bar max. differential pressure (30 psi) at 25 °C (Reverse; intermittent)
25 °C max. inlet temp.
5.5 bar max. inlet pressure (80 psi) at 23 °C
5500 mbar max. differential pressure (80 psi) at 25 °C (Forward)
6.9 bar max. inlet pressure (100 psi) at 23 °C (intermittent)
6900 mbar max. differential pressure (100 psi) at 25 °C (Forward; intermittent)
80 psig max. inlet pressure
técnica(s)
sterile filtration: suitable
Largura
7.8 in.
comprimento nominal do cartucho
30 in. (75 cm)
device L
87.2 cm (34.3 in.)
device size
30 in.
diâmetro do filtro
19.8 cm (7.8 in.)
filter filtration area
1.71 m2
Impurezas
≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.25 EU/mL USP bacterial endotoxins (sample aqueous extraction)
matriz
Millipore Express® SHF
tamanho de poro
0.2 μm pore size
ponto de bolha
≥4000 mbar (58 psig), air with water
conector
(25 mm (1 in.) Hose Barb Inlet and Outlet)
Procurando produtos similares? Visita Guia de comparação de produtos
Descrição geral
Embalagem
Nota de preparo
3 autoclave cycles of 60 min at 126 °C; capable of 45 kilogray (4.5 Megarad) gamma exposure; not in-line steam sterilizable
Nota de análise
Outras notas
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer installation section of OPTICAP® XL Capsules and OPTICAP® XLT Capsules user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations
Informações legais
Não está encontrando o produto certo?
Experimente o nosso Ferramenta de seleção de produtos.
Certificados de análise (COA)
Busque Certificados de análise (COA) digitando o Número do Lote do produto. Os números de lote e remessa podem ser encontrados no rótulo de um produto após a palavra “Lot” ou “Batch”.
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica