749-0798
Semivolatiles dans l′eau
suitable for GC, application for SPE
About This Item
Produits recommandés
Technique(s)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
Paramètres du test
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: acidified drinking water
SPE tube/cartridge: ENVI-18 DSK SPE Disk, 47 mm (57171)
condition: 5 mL dichloromethane; 5 mL methanol; 5 mL DI water
sample addition: 5 mL water sample
elution: 2 x 5 mL acetonitrile rinsing internal chamber of sample reservoir thoroughly
column: 5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven: 40 °C to 160 °C (3 min), then to 300 °C (3 min) at 6 °C/min
detector: MS, scan range m/z = 45-450
carrier gas: helium, 33 cm/sec
injection: 1 μL split/splitless, 1 min delay
Adéquation
application for SPE
Application(s)
environmental
food and beverages
Application
Ajuster 1L d′eau potable à pH<2 avec de l′HCl 6N. Ajouter 5mL de MeOH et mélanger énergiquement
Conditionner le disque de SPE avec 5mL de dichlorométhane (pousser DCM complètement à travers le disque) et ajouter 5mL de MeOH
Equilibreravec 5mL d′H2O DI
Charger l′échantillon d′eau, 100mL/min
Eluer avec 2 x 5mL de MeCN en rinçant avec précaussion la chambre interne du réservoir d′échantillon
Informations légales
Analyte
Colonne analytique
Produit(s) apparenté(s)
Choose from one of the most recent versions:
Certificats d'analyse (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique