HPLC Analysis of Sugars and Taurine in an Energy Drink on Ascentis® Express HILIC using UV and ELSD Detection in Series

HPLC Analysis of Sugars and Taurine in an Energy Drink on Ascentis® Express HILIC using UV and ELSD Detection in Series application for HPLC

Materialien

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Ascentis® Express HILIC HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 3 mm

Standard

Produkt-Nr.
Beschreibung
Preisangaben

Saccharose

≥99.5% (GC)

Taurin

≥99%

Fructose

United States Pharmacopeia (USP) Reference Standard

Glucose

European Pharmacopoeia (EP) Reference Standard

CONDITIONS

column

Ascentis Express HILlC, 10 cm x 3.0 mm I.D., 2.7 μm particles (53970-U)

mobile phase

[A] 100 mM ammonium acetate, pH 5.0 with acetic acid; [B] water; [C] acetonitrile; (09:01:90, A:B:C)

flow rate

0.6 mL/min

pressure

812 psi (56 bar)

column temp.

35 °C

detector

ELSD, 55 °C, 3.5 bar nitrogen

injection

2 μL

sample

dilute1:9 in acetonitrile

Beschreibung

Hinweis zur Analyse

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of common drink additives used in caffeinated energy drinks. The analysis was performed with UV and ELSD detectors in series, this chromatogram being from the ELSD.