Stericup® and Steritop® filters are fitted with a 1/4” tube adapter. We recommend 3/16 in. (4.8 mm) ID silicone tubing to fit this tube adapter: https://www.sigmaaldrich.com/product/mm/xx7100004
Größe auswählen
| Größe/SKU | Verfügbarkeit | Preis |
|---|
Über diesen Artikel
Quality Segment
material
polyethersulfone membrane, polystyrene
product line
Steritop®
feature
holdup volume 2.4 mL, hydrophilic
packaging
pack of 12 ea
manufacturer/tradename
Steritop®
parameter
250 mL process volume, 250 mL sample volume, 45 °C max. temp.
technique(s)
sterile filtration: suitable
H
95 mm
diam.
12.5 cm
filter capacity
250 mL
filter diam.
33 mm
filtration area
40 cm2
matrix
Express PLUS™
pore size
0.22 μm nominal pore size, 0.22 μm pore size
fitting
inlet funnel, 33 mm outlet thread(s)
shipped in
ambient
storage temp.
15-25°C
General description
Die Stericup ® und Steritop ® Systeme für die Sterilfiltration verwenden zuverlässige Hochleistungsmembranen für eine optimale Sterilfiltration von Laborreagenzien, Puffern und Medien. Diese Einheiten bieten hervorragende Durchfluss- und Durchsatzraten mit geringer unspezifischer Bindung und ein stabiles Design mit Kippschutz.
Dieser Steritop ® 250-ml-Flaschenaufsatz mit einer Membranfiltereinheit aus Polyethersulfon (PES) mit einer Porengröße von 0,22 μm kann auf Flaschen mit 33-mm-Öffnungen aufgesetzt werden und ist für Vakuumanwendungen bestimmt.
Application
- Gewebekulturmedien +/- Additive
- Puffer
- Biologische Lösungen
Legal Information
1 of 1
Dieser Artikel | |||
|---|---|---|---|
| description pore size 0.22 μm, polyethersulfone membrane, filter capacity 250 mL, 33 mm outlet thread(s), pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 150 mL, 33 mm outlet thread(s), pack of 12 ea | description pore size 0.1 μm, polyethersulfone membrane, filter capacity 1000 mL, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 500 mL, 33 mm outlet thread(s), pack of 12 ea |
| manufacturer/tradename Steritop® | manufacturer/tradename - | manufacturer/tradename - | manufacturer/tradename - |
| filter capacity 250 mL | filter capacity 150 mL | filter capacity 1000 mL | filter capacity 500 mL |
| pore size 0.22 μm pore size, 0.22 μm nominal pore size | pore size 0.22 μm pore size | pore size 0.1 μm pore size | pore size 0.22 μm pore size |
| material polystyrene , polyethersulfone membrane | material polystyrene funnel(s), polystyrene housing, polyethersulfone membrane | material polyethersulfone membrane, polystyrene bottle, polystyrene housing, polystyrene funnel(s) | material polyethersulfone membrane, polystyrene housing, polystyrene funnel(s) |
| fitting inlet funnel, 33 mm outlet thread(s) | fitting inlet funnel, 33 mm outlet thread(s) | fitting 45 mm outlet thread(s), inlet funnel | fitting 33 mm outlet thread(s), inlet funnel |
| feature holdup volume 2.4 mL, hydrophilic | feature hydrophilic | feature hydrophilic | feature hydrophilic |
Still not finding the right product?
Explore all of our products under Millipore® Steritop® Vakuumflaschenaufsatzfilter
Hier finden Sie alle aktuellen Versionen:
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
-
What tubing can be used with Stericup® and Steritop® vacuum filters?
What tubing can be used with Stericup® and Steritop® vacuum filters?
1 answer-
Helpful?
-
-
What is the maximum vacuum pressure for filtration?
1 answer-
700 mm Hg (27.5” Hg, 0.9 bar) differential vacuum at 25 °C (77 °F)
Helpful?
-
-
What is the maximum processing volume for Stericup® and Steritop® filters? Can I add more volume after initial filtration?
1 answer-
All Stericup® and Steritop® filters were developed for single-use application. The maximum volume is dependent on the receiver bottle capacity. Stericup® filters come with a receiver bottle with a specified maximum volume. For Steritop® filters, the maximum recommended volume is 2 L with an appropriate-sized receiver bottle. Larger processing volumes can result in membrane fouling, increase chance of spillage and contamination, and lead to possible airlock*.
*Note that devices with PES membranes can airlock if filtered to dryness. Some liquid must be left on top of the filter, then the system stopped, receiver changed or emptied, and more sample volume added before restart. This increases the chance of spillage.Helpful?
-
-
What leachables are detected during filtration processes using Stericup® filters?
1 answer-
We have not performed extractables/leachables studies for our Stericup® filters.
Helpful?
-
-
Are Stericup® and Steritop® filters GMP-certified?
1 answer-
No, Stericup® and Steritop® devices are not GMP-approved.
Helpful?
-
-
The User Guide says “Use only glass or plastic bottles designed for vacuum applications”. Can you specify which bottles are compatible?
1 answer-
Steritop® Quick Release bottle top units require a vacuum-rated receiver bottle with either a GL33 or GL45 mm closure thread as specified. Steritop® E bottle top units require a vacuum-rated receiver bottle with a GL45 mm closure thread. Bottles no larger than 2 L should be used.
Glass bottles, GL33:
• Wheaton media bottles (glass), 500 mL (Product Z364886)
• Wheaton media bottles (glass), 1 L (Product Z364894)
Glass bottles, GL45
• Corning® glass media storage bottle, 125 mL (Product CLS1399125)
• Corning® glass media storage bottle, 250 mL (Product CLS1399250)
• Corning® glass media storage bottle, 500 mL (Product CLS1399500)
• Corning® glass media storage bottle, 1 L (Product CLS13991L)
• Pyrex® round media storage bottle, 500 mL (Product CLS1395500)
• Pyrex® round media storage bottle, 1 L (Product CLS13951L)
• KIMBLE® KIMAX® Media bottle, 250 mL (Product DWK14395250)
• KIMBLE® KIMAX® Media bottle, 500 mL (Product DWK14395500)
• KIMBLE® KIMAX® Media bottle, 1 L (Product DWK143951000)
Plastic (pre-sterile) bottles, GL45:
• Stericup® Quick Release Receiver flask, 250 mL (Product S200B02RE)
• Stericup® Quick Release Receiver flask, 500 mL (Product S200B05RE)
• Stericup® Quick Release Receiver flask, 1000 mL (Product S200B10RE)Helpful?
-
-
Which type of pump should be used with Stericup® and Steritop® vacuum filters?
Which type of pump should be used with Stericup® and Steritop® vacuum filters?
1 answer-
We recommend our Chemical Duty Pump: https://www.sigmaaldrich.com/product/mm/wp6111560
Helpful?
-
-
What is the chemical compatibility of the various Stericup® and Steritop® filters?
What is the chemical compatibility of the various Stericup® and Steritop® filters?
1 answer-
Please refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
Helpful?
-
-
Can Stericup® and Steritop® filters be reused?
1 answer-
All Stericup® and Steritop® filters were developed for single use, disposable processing.
Helpful?
-
-
Can we filter oil-type liquids with Stericup® filters?
Can we filter oil-type liquids with Stericup® filters?
1 answer-
Likely not, due to pore size/possible fouling/chemical compatibility. Refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
Helpful?
-