G005284
GC Analysis of Glycols in Water on SPB®-1000 after SPE using ENVI-Carb™ Plus
suitable for GC, application for SPE
About This Item
Productos recomendados
technique(s)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: tap water spiked at 2 μg/mL with glycols
SPE tube/cartridge: ENVI-Carb Plus reversible cartridge, 400 mg/ 1mL (54812-U)
condition: 1 mL methylene chloride, 2 x 2 mL methanol, 3 mL deionized water (cartridge not allowed to dry during conditioning)
sample addition: 5 mL spiked tap water
drying: 10 min, 10 mm mercury
elution: 2 mL 80:20 methanol:methylene chloride
column: SPB-1000, 30 m x 0.25 mm I.D., 0.25 μm (24313)
oven: 50 °C (1 min), 8 °C/min to 200 °C (7 min)
inj. temp.: 220 °C
detector: FID, 220 °C
carrier gas: helium, 1.5 mL/min. constant
injection: 1 μL, splitless (splitter open at 0.75 min.)
liner: 4 mm I.D. FocusLiner™ inlet liner with taper
suitability
application for SPE
application(s)
environmental
Legal Information
Analito
SPE tube or plate
analytical column
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico