技術
gas chromatography (GC): suitable
測試參數
column: SPB-608, 30 m x 0.25 mm I.D., 0.25 μm (24103-U)
oven: 70 °C (0.2 min), 25 °C/min to 265 °C, 5 °C/min to 300 °C (12 min)
inj. temp.: 250 °C
detector: FlD, 300 °C
carrier gas: helium, 1.3 mL/min, constant
injection: 2.0 μL pulsed splitless (30 psi until 0.2 min)
liner: 4 mm I.D. single taper
sample: PAH standard, 1 μg/mL in methylene chloride
應用
environmental
分析報告
US EPA Method 610 describes the analysis of polynuclear aromatic hydrocarbons (commonIy referred to as PAHs or PNAs) by both HPLC and GC. The method designates the use of a packed GC column, however capillary columns will provide better resolution, baseline stability and longer lifetime. This application illustrates the separation of the target compound list in this method on the SPB-608 capillary GC column. Overall analysis time on the SPB-608 is longer than the SLB-5ms due to stronger retention and its lower maximum operating temperature. A pressure pulsed injection was used to improve response of the heavier PAHs, and to reduce the larger vapor volume resulting from a 2 μL injection.
法律資訊
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
描述
1-Fluoronaphthalene
Naphthalene
2-Fluorobiphenyl
Acenaphthylene
Acenaphthene
Fluorene
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Chrysene
Benzo[b]fluoranthene
Benzo[k]fluoranthene
Benzo[a]pyrene
Indeno[1,2,3-cd]pyrene
Benzo[ghi]perylene
查看全部 (16)
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務