技術
HPLC: suitable
測試參數
column: Ascentis Express C8, 5 cm x 3 mm I.D., 2.7 μm particles (53848-U)
mobile phase: [A] 0.1% formic acid in water; [B] 0.1% formic acid in acetonitrile
gradient: 15 to 100% B in 8.5 min
flow rate: 0.6 mL/min
column temp.: 35 °C
detector: UV, 214 nm
injection: 2 μL
sample: 100 mg/L in 75 μM EDTA, 1.5 mM ascorbate, 15% acetonitrile
適合性
application for HPLC
應用
vitamins, nutraceuticals, and natural products
分析報告
This application demonstrates the suitability of Ascentis Express C8 for the efficient separation of phenolics from rosemary.
法律資訊
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
描述
Rosmarinic acid
Carnosol
Carnosic acid
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務