技術
HPLC: suitable
測試參數
column: Ascentis Express Phenyl-Hexyl, 5 cm x 4.6 mm I.D., 2.7 μm particles (53348-U)
mobile phase: [A] water; [B] methanol; (35:65, A:B)
flow rate: 1.4 mL/min
pressure: 3600 psi (248 bar)
column temp.: 35 °C
detector: UV, 250 nm
injection: 2 μL
sample: 0.1 g/L in 20% methanol
適合性
application for HPLC
應用
clinical
pharmaceutical (small molecule)
分析報告
The Phenyl-Hexyl phase of the Ascentis Express product line displays the best selectivity of this sample; selectivity is markedly affected by choice of organic modifier. The application also demonstrates feasibility of fast LC with Ascentis Express by virtue of low backpressure and high efficiency.
法律資訊
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
描述
Clonazepam
Oxazepam
Alprazolam
Desmethyldiazepam
Diazepam
分析柱
標準
產品號碼
描述
訂價
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務