跳轉至內容
Merck
全部照片(1)

文件

G004415

Improved HPLC Selectivity of Discovery® BIO Wide Pore RP Phases

application for HPLC

登入查看組織和合約定價


About This Item

技術

HPLC: suitable

測試參數

column: Discovery BIO Wide Pore C18, 15 cm x 4.6 mm I.D., 5 μm particles vs. competitive wide pore C18 column of same dimensions (568222-U)
mobile phase: [A] 0.1% TFA in 80:20 water:acetonitrile; [B] 0.1% TFA in 66:34 water:acetonitrile

Gradient of 0 to 100% B in 14 min. after 1 min. delay

gradient: 0 to 100% B in 14 min. after 1 min. delay
flow rate: 1 mL/min
column temp.: 30 °C
detector: UV, 220 nm
injection: 10 μL
sample: peptide mixture (Sigma P2693), ~0.25 μg each peptide in 0.1% TFA

適合性

application for HPLC

應用

life science and biopharma

法律資訊

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany

Choose from one of the most recent versions:

分析證明 (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務