跳轉至內容
Merck
全部照片(1)

重要文件

797-0049

HPLC Analysis of Carbamate and Urea Pesticides on SUPELCOSIL LC-8 after SPME using 60 μm PDMS/DVB Fiber

application for SPME, application for HPLC

登入查看組織和合約定價


About This Item

暫時無法取得訂價和供貨情況

技術

HPLC: suitable
solid phase microextraction (SPME): suitable

測試參數

sample/matrix: 3 mL water containing 8ng/mL of each analyte in 10% NaCl
SPME fiber: PDMS/DVB, 60 μm (57317)
extraction: immersion, 40 min, rapid stirring
desorption process: static, 5 min in acetonitrile:water (65:35); dynamic, valve open during run
column: SUPELCOSIL LC-8, 15 cm × 4.6 mm I.D., 5 μm particles (58220-U)
mobile phase: [A] acetonitrile; [B] water
gradient: 18:82 to 65:35 in 9 min, hold 3 min
flow rate: 2.0 mL/min
column temp.: 35 °C
detector: UV, 240 nm

適合性

application for HPLC
application for SPME

應用

environmental

法律資訊

SUPELCOSIL is a trademark of Sigma-Aldrich Co. LLC

分析物

描述

    Propham, analytical standard
    Siduron, ampule of 1000 mg
    Linuron, analytical standard
    Chlorpropham, analytical standard
    Barban, analytical standard
    Neburon, PESTANAL®, analytical standard

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 文件 section.

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務