跳轉至內容
Merck
全部照片(1)

重要文件

794-0863

GC Analysis of Gasoline Residue in Arson Samples on a 100% Methyl Column after SPME using 100 μm PDMS Fiber

suitable for GC, application for SPME

登入查看組織和合約定價


About This Item

暫時無法取得訂價和供貨情況

技術

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

測試參數

SPME fiber: polydimethylsiloxane phase fiber, 100 μm headspace sampling (20 min) 10 sec desorption (splitless mode) (57300-U)
column: 100% methyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, Equity-1, 28046-U)
oven: 35 °C (2 min) to 220 °C at 10 °C/min, hold 2 min, to 300 °C at 30 °C/min, hold 5 min
inj. temp.: splitless (closed 3 min), 220 °C (2 mm I.D. injector liner)
detector: FID, 300 °C
carrier gas: helium, 1 mL/min (split 50:1)

適合性

application for SPME

應用

forensics and toxicology

分析物

描述

    m-Xylene
    p-Xylene
    Naphthalene
    Toluene
    Ethylbenzene
    o-Xylene
    1,2,3-Trimethylbenzene
    1,2,4-Trimethylbenzene
    2-Methylnaphthalene
    1-Methylnaphthalene
    查看全部 (10)

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 文件 section.

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務