As listed in the Properties section this is a Durapore membrane.
S2GVU05RE
Millipore® Stericup®Quick Release 真空过滤系统
pore size 0.22 μm, PVDF membrane, funnel capacity 500 mL, pack of 12 ea
别名:
SCGVU05RE, Stericup Quick Release-GV无菌真空过滤系统
选择尺寸
About This Item
推荐产品
材料
PVDF membrane
polystyrene bottle
polystyrene funnel(s)
polystyrene housing
品質等級
無菌
sterile; irradiated
產品線
Stericup®
特點
hydrophilic
包裝
pack of 12 ea
參數
max. temp. at 45 °C
描述
2.4 mL (hold-up volume)
技術
sterile filtration: suitable
高度
263 mm
容量
500 mL
直徑
12.5 cm
過濾器直徑
73 mm
過濾面積
40 cm2
漏斗容量
500 mL
基質
Durapore®
孔徑
0.22 μm pore size
接頭
inlet funnel
outlet flask (Receiver)
運輸包裝
ambient
一般說明
Stericup ®和Steritop ®无菌过滤器设备采用可靠的高性能膜, 可实现实验室试剂、缓冲液和培养基的最佳无菌过滤操作 。 这些设备将良好的流速和处理量与低非特异性结合和稳定的无尖端设计 结合起来。
这种Stericup® 0.22 μm 孔径偏氟乙烯(PES)膜过滤器装置 结合了500 mL 瓶顶过滤器和500 mL接收瓶及瓶盖, 便于加工和储存。
應用
- 组织培养基 +/- 添加剂
- 缓冲液
- 生物溶液
特點和優勢
- 该 Quick Release 过滤器漏斗旋转四分之一圈即可从接收瓶上拆下 —减少溢出的可能性,并且 在实验台或通风橱中更容易操作。
- Steritop ®过滤器漏斗上的Bold产品标识清楚地 显示了膜特性,包括组成和孔径, 可用于快速选择适合应用的产品。
- 密封盖配有法兰并进行改良以便在潮湿和干燥条件下均能握紧 —可触摸塞确保牢固的密封。不用担心 灭菌的内容物可能受污染或溢出
- 接收瓶上充足的磨砂书写表面和更浅的瓶盖颜色 便于清楚标识,以改善易读性并减少不可读的 污迹。
- 密封盖位于侧面以防污染风险。
- 接收瓶可堆叠以节省储存空间。
法律資訊
商品
Learn about the impact of filtration on hematopoietic cell culture media containing vital T cell factors and other media components.
实验方案
SEC-MS protocol for rapid glycoprofiling of monoclonal antibodies with antibody purification and mass spectrometer calibration.
-
Is this PVDF membrane a durapore membrane?
1 answer-
Helpful?
-
-
What is the difference between classic Stericup® filters and Stericup® “Quick Release” filters?
1 answer-
Stericup® Quick Release, or “QR”, filters feature a connection between funnel filter and receiver bottle that requires only a quarter-turn release to facilitate funnel removal. These filters come fully engaged – there is no need to tighten before use. The receiver bottle cap has flanges on the top surface to improve grip, as well as a positive lock stop to ensure a leak-proof seal with the receiver bottle. Other improvements include a large, frosted writing surface on the receiver bottle and a lighter cap color to facilitate labelling, more prominent volume markings on the filtration funnel and receiver bottle, and a product label with a peel-off lot record strip on the bag for easy viewing in storage.
Helpful?
-
-
Can Stericup® bottles be placed inside freezers for long-term storage of media?
1 answer-
You can successfully freeze and store many aqueous solutions (such as culture media) in Stericup® bottles at temperatures up to -20 °C (-4 °F). It is strongly recommended that you run a sample stability trial under your actual storage conditions prior to using Stericup® bottles for frozen storage.
Helpful?
-
-
What leachables are detected during filtration processes using Stericup® filters?
1 answer-
We have not performed extractables/leachables studies for our Stericup® filters.
Helpful?
-
-
What Bacterial Endotoxins Test is used?
1 answer-
The test method used complies with USP 85 and EP 2.6.14 to ensure the absence of endotoxin in our products.
Helpful?
-
-
What is the expiration date of Stericup® and Steritop® filters?
1 answer-
Typically 3 years from sterilization.
Helpful?
-
-
What is the chemical compatibility of the various Stericup® and Steritop® filters?
What is the chemical compatibility of the various Stericup® and Steritop® filters?
1 answer-
Please refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
Helpful?
-
-
What tubing can be used with Stericup® and Steritop® vacuum filters?
What tubing can be used with Stericup® and Steritop® vacuum filters?
1 answer-
Stericup® and Steritop® filters are fitted with a 1/4” tube adapter. We recommend 3/16 in. (4.8 mm) ID silicone tubing to fit this tube adapter: https://www.sigmaaldrich.com/product/mm/xx7100004
Helpful?
-
-
Are Stericup® and Steritop® filters GMP-certified?
1 answer-
No, Stericup® and Steritop® devices are not GMP-approved.
Helpful?
-
-
What is the maximum processing volume for Stericup® and Steritop® filters? Can I add more volume after initial filtration?
1 answer-
All Stericup® and Steritop® filters were developed for single-use application. The maximum volume is dependent on the receiver bottle capacity. Stericup® filters come with a receiver bottle with a specified maximum volume. For Steritop® filters, the maximum recommended volume is 2 L with an appropriate-sized receiver bottle. Larger processing volumes can result in membrane fouling, increase chance of spillage and contamination, and lead to possible airlock*.
*Note that devices with PES membranes can airlock if filtered to dryness. Some liquid must be left on top of the filter, then the system stopped, receiver changed or emptied, and more sample volume added before restart. This increases the chance of spillage.Helpful?
-
Active Filters
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门