GC Analysis of Basic Drugs on SLB®-5ms (20 m x 0.18 mm I.D., 0.36 μm) after SPME using 85 μm Polyacrylate Fiber, Fast GC Analysis

suitable for GC, application for SPME

Zaloguj sięWyświetlanie cen organizacyjnych i kontraktowych


About This Item

metody

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

parametry testowe

sample/matrix: 500 ug/L of each compound in 2 mL 0.05M phosphate buffer, pH=11, plus sodium chloride (NaCl) to 25% (w/v)
SPME fiber: 85 μm Polyacrylate (57304)
extraction: immersion, 450 rpm stirring at room temp. for 15 min
desorption process: 250 °C for 3 min
column: SLB-5ms, 20 m x 0.18 mm I.D. x 0.36 μm (28576-U)
oven: 100 °C (1 min), 20 °C/min to 325 °C (5 min)
MSD interface: 325 °C
carrier gas: helium, 1.5 mL/min. constant
liner: 0.75 mm I.D. SPME

przydatność

application for SPME

Zastosowanie

forensics and toxicology

Informacje prawne

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
This page may contain text that has been machine translated.

Choose from one of the most recent versions:

Certyfikaty analizy (CoA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Obsługa Klienta

Masz już ten produkt?

Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.

Odwiedź Bibliotekę dokumentów

Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.

Skontaktuj się z zespołem ds. pomocy technicznej