GC Analysis of PCBs in Transformer Oil on a 5% Phenyl Column after SPE using Supelclean™ Sulfoxide
suitable for GC, application for SPE
About This Item
metody
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
parametry testowe
SPE tube/cartridge: Supelclean Sulfoxide SPE Tube, Glass 6 g/20 mL (55252-U)
sample preparation: SPE (Solid Phase Extraction)
column: 5% phenyl polysiloxane (Supelco equivalent, SLB-5ms, 28471-U)
sample addition: transformer oil spiked with PCBs
detector: QMS (For full method details, please see reference (1))
extraction process: Sample Pre-treatment: paraffin oil was spiked with a Kanechlor PCB mix at the total levels of 3.7 ppm (mg/kg) and diluted with hexane (1:1 v/v). (Commercial insulation oil (Japan Industrial Standard JIS C2320-1999, insulating oil, Class 1-2/4).)
condition: 20 mL acetone; 40 mL of hexane
sample addition: 0.4 mL diluted oil sample
elution: 12 mL hexane (elutes oil interferences); 25 mL hexane (elutes PCBs)
eluate post-treatment: concentrate under nitrogen
przydatność
application for SPE
Zastosowanie
environmental
Zastosowanie
Commercial insulation oil (Japan Industrial Standard JIS C2320-1999, insulating oil, Class 1-2/4, paraffin oil) was spiked with a Kanechlor PCB mix at the total levels of 3.7 ppm (mg/kg) and diluted with hexane (1:1 v/v).
Informacje prawne
kolumna analityczna
probówka lub płytka SPE
Wybierz jedną z najnowszych wersji:
Certyfikaty analizy (CoA)
Przepraszamy, ale COA dla tego produktu nie jest aktualnie dostępny online.
Proszę o kontakt, jeśli potrzebna jest pomoc Obsługa Klienta
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej