All Stericup® and Steritop® filters were developed for single use, disposable processing.
SCGPS02RE
Millipore® Steritop® Vacuum Bottle Top Filter
pore size 0.22 μm, polyethersulfone membrane, funnel capacity 250 mL, 33 mm outlet thread(s), pack of 12 ea
동의어(들):
Steritop Threaded Bottle Top Filter, Steritop®-GP Quick Release Bottle Top Filter, bottle top sterile filter unit, bottle top vacuum filter, bottle top vacuum filtering unit
About This Item
추천 제품
material
polyethersulfone membrane
polystyrene
Quality Level
무균
sterile
제품 라인
Steritop®
특징
holdup volume 2.4 mL
hydrophilic
포장
pack of 12 ea
제조업체/상표
Steritop®
파라미터
250 mL process volume
250 mL sample volume
45 °C max. temp.
기술
sterile filtration: suitable
높이
95 mm
직경
12.5 cm
필터 직경
33 mm
여과 면적
40 cm2
깔대기 용량
250 mL
Matrix
Express PLUS™
공극 크기
0.22 μm nominal pore size
0.22 μm pore size
피팅
inlet funnel
33 mm outlet thread(s)
배송 상태
ambient
일반 설명
Stericup ® and Steritop ® sterile filtration devices use reliable, high-performance membranes for optimum sterile filtration of laboratory reagents, buffers, and media. These devices combine superior flow rates and throughput with low non-specific binding and a stable, no-tip design.
This Steritop ® 250 mL bottle top 0.22 μm pore size polyethersulfone (PES) membrane filter unit can be used on bottles with 33 mm openings, , designed for vacuum applications.
애플리케이션
- Tissue culture media +/- additives
- Buffers
- Biological solutions
법적 정보
시험 성적서(COA)
제품의 로트/배치 번호를 입력하여 시험 성적서(COA)을 검색하십시오. 로트 및 배치 번호는 제품 라벨에 있는 ‘로트’ 또는 ‘배치’라는 용어 뒤에서 찾을 수 있습니다.
이미 열람한 고객
관련 콘텐츠
What are you filtering out? What do you want to keep? Pick the right membrane for your filtration.
This article shows that a PES membrane containing Stericup Quick Release filter provides an alternative for the pluripotent expansion of embryonic stem cells.
-
Can Stericup® and Steritop® filters be reused?
1 답변-
도움이 되었습니까?
-
-
What is the difference between classic Stericup® filters and Stericup® “Quick Release” filters?
1 답변-
Stericup® Quick Release, or “QR”, filters feature a connection between funnel filter and receiver bottle that requires only a quarter-turn release to facilitate funnel removal. These filters come fully engaged – there is no need to tighten before use. The receiver bottle cap has flanges on the top surface to improve grip, as well as a positive lock stop to ensure a leak-proof seal with the receiver bottle. Other improvements include a large, frosted writing surface on the receiver bottle and a lighter cap color to facilitate labelling, more prominent volume markings on the filtration funnel and receiver bottle, and a product label with a peel-off lot record strip on the bag for easy viewing in storage.
도움이 되었습니까?
-
-
Can we filter oil-type liquids with Stericup® filters?
Can we filter oil-type liquids with Stericup® filters?
1 답변-
Likely not, due to pore size/possible fouling/chemical compatibility. Refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
도움이 되었습니까?
-
-
Are Stericup® and Steritop® filters GMP-certified?
1 답변-
No, Stericup® and Steritop® devices are not GMP-approved.
도움이 되었습니까?
-
-
The User Guide says “Use only glass or plastic bottles designed for vacuum applications”. Can you specify which bottles are compatible?
1 답변-
Steritop® Quick Release bottle top units require a vacuum-rated receiver bottle with either a GL33 or GL45 mm closure thread as specified. Steritop® E bottle top units require a vacuum-rated receiver bottle with a GL45 mm closure thread. Bottles no larger than 2 L should be used.
Glass bottles, GL33:
• Wheaton media bottles (glass), 500 mL (Product Z364886)
• Wheaton media bottles (glass), 1 L (Product Z364894)
Glass bottles, GL45
• Corning® glass media storage bottle, 125 mL (Product CLS1399125)
• Corning® glass media storage bottle, 250 mL (Product CLS1399250)
• Corning® glass media storage bottle, 500 mL (Product CLS1399500)
• Corning® glass media storage bottle, 1 L (Product CLS13991L)
• Pyrex® round media storage bottle, 500 mL (Product CLS1395500)
• Pyrex® round media storage bottle, 1 L (Product CLS13951L)
• KIMBLE® KIMAX® Media bottle, 250 mL (Product DWK14395250)
• KIMBLE® KIMAX® Media bottle, 500 mL (Product DWK14395500)
• KIMBLE® KIMAX® Media bottle, 1 L (Product DWK143951000)
Plastic (pre-sterile) bottles, GL45:
• Stericup® Quick Release Receiver flask, 250 mL (Product S200B02RE)
• Stericup® Quick Release Receiver flask, 500 mL (Product S200B05RE)
• Stericup® Quick Release Receiver flask, 1000 mL (Product S200B10RE)도움이 되었습니까?
-
-
Which type of pump should be used with Stericup® and Steritop® vacuum filters?
Which type of pump should be used with Stericup® and Steritop® vacuum filters?
1 답변-
We recommend our Chemical Duty Pump: https://www.sigmaaldrich.com/product/mm/wp6111560
도움이 되었습니까?
-
-
What is the maximum processing volume for Stericup® and Steritop® filters? Can I add more volume after initial filtration?
1 답변-
All Stericup® and Steritop® filters were developed for single-use application. The maximum volume is dependent on the receiver bottle capacity. Stericup® filters come with a receiver bottle with a specified maximum volume. For Steritop® filters, the maximum recommended volume is 2 L with an appropriate-sized receiver bottle. Larger processing volumes can result in membrane fouling, increase chance of spillage and contamination, and lead to possible airlock*.
*Note that devices with PES membranes can airlock if filtered to dryness. Some liquid must be left on top of the filter, then the system stopped, receiver changed or emptied, and more sample volume added before restart. This increases the chance of spillage.도움이 되었습니까?
-
-
What tubing can be used with Stericup® and Steritop® vacuum filters?
What tubing can be used with Stericup® and Steritop® vacuum filters?
1 답변-
Stericup® and Steritop® filters are fitted with a 1/4” tube adapter. We recommend 3/16 in. (4.8 mm) ID silicone tubing to fit this tube adapter: https://www.sigmaaldrich.com/product/mm/xx7100004
도움이 되었습니까?
-
-
What is the maximum vacuum pressure for filtration?
1 답변-
700 mm Hg (27.5” Hg, 0.9 bar) differential vacuum at 25 °C (77 °F)
도움이 되었습니까?
-
-
What is the chemical compatibility of the various Stericup® and Steritop® filters?
What is the chemical compatibility of the various Stericup® and Steritop® filters?
1 답변-
Please refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
도움이 되었습니까?
-