コンテンツへスキップ
Merck

HPLC Analysis of B-Vitamins in Vitamin Water using Ascentis® Express RP-Amide and UV Detection

HPLC Analysis of B-Vitamins in Vitamin Water using Ascentis® Express RP-Amide and UV Detection application for UHPLC

材料

分析カラム

製品番号
詳細
価格

Ascentis® Express RP-アミド, 2.7 μm HPLCカラム

2.7 μm particle size, L × I.D. 10 cm × 3 mm

CONDITIONS

column

Ascentis Express RP-Amide, 10 cm x 3.0 mm I.D., 2.7 μm particles (53918-U)

mobile phase

[A] 20 mM potassium phosphate, pH 3; [B] methanol

gradient

held at 0.5% B for 1.5 minutes, to 25% B in 1.0 minutes, held at 25% B for 1.7 minute, held at 0.5% B for 1.2 minutes

flow rate

0.75 mL/min

pressure

5584 psi (385 bar)

column temp.

30 °C

detector

UV, 210 nm

injection

0.5 μL

sample

vitamin water was filtered through 0.2um nylon syringe filter

詳細

アナリシスノート

The vitamin water drink was filtered using 0.2 um Nylon filter and injected into UHPLC for B-vitamin analysis

法的情報