UHPLC/MS Analysis of Diazepam and Metabolites on Titan C18, Comparison of Flow Rate and Column Length
材料
標準
製品番号
詳細
価格
CONDITIONS
column
Titan C18, 2 cm x 2.1 mm I.D., 1.9 μm particles (577120-U)
mobile phase
[A] 0.1% formic acid; [B] acetonitrile; (60:40, A:B)
flow rate
various
temp.
35 °C
detector
TIC MRM (UV also runs in series)
injection
1 μL
sample
100 ng/mL in mobile phase
詳細
アナリシスノート
These two sets of chromatograms show separations on 100, 50, 30, and 20 mm x 2.1 mm I.D. columns and at different flow rates and demonstrate the utility of the 1.9 μm Titan particles. The highest grade solvents provided clean, robust operation. Cerilliant CRMs provided reliable quantification.
法的情報
null