LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 4

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 4 application for HPLC

材料

アプリケーション

製品番号
詳細
価格

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 1

application for HPLC

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 2

application for HPLC

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 3

application for HPLC

分析カラム

製品番号
詳細
価格

Ascentis® Express HILIC、2.7 μm HPLCカラム

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

標準

製品番号
詳細
価格

Mephedrone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Methylone hydrochloride

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

ピロバレロン 塩酸塩 溶液

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

N-Ethylcathinone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)

mobile phase

5 mM ammonium formate in [A] acetonitrile; [B] water; (98:2, A:B)

flow rate

0.6 mL/min

pressure

267 psi (18.4 bar)

column temp.

35 °C

detector

MS, ESI(+) TOF, Summed Ion Chromatogram

injection

2 μL

sample

500 ng/mL in acetonitrile

詳細

アナリシスノート

HILIC mode operation on an Ascentis Express column provided baseline resolution of these bath salts, including isobaric compounds. Ultra high purity solvents were used to supply low background interference and low particulate contaminants for robust; trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.