コンテンツへスキップ
Merck

HPLC Analysis of Protein Standards on Zenix® SEC-300 by Size Exclusion As Affected by Arginine

HPLC Analysis of Protein Standards on Zenix® SEC-300 by Size Exclusion As Affected by Arginine application for HPLC

材料

付属品

製品番号
詳細
価格

EXP®Pre-カラムフィルター

holder with EXP titanium hybrid ferrule, pack of 1 sets

EXP® Pre-Column Filter Cartridges

0.5μm, pkg of 5 ea

分析カラム

製品番号
詳細
価格

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 4.6 mm, 3 μm particle size, stainless steel

移動相成分

製品番号
詳細
価格

L-アルギニン

from non-animal source, meets EP, USP testing specifications, suitable for cell culture, 98.5-101.0%

CONDITIONS

column

Zenix SEC-300, 30 cm x 4.6 mm I.D., 3 μm particles (A777028)

mobile phase

[A], 0.15 M potassium phosphate, monobasic, pH 7.0 with potassium hydroxide; [B] 0.15 M potassium phosphate, monobasic, 0.2 M arginine, pH 7.0 with phosphoric acid (mobile phase is either 100% A or 100% B)

flow rate

0.3 mL/min

pressure

1030 psi 100% A (71 bar)

pressure

1165 psi 100% B (80 bar)

column temp.

25 °C

detector

UV, 280 nm

injection

2.5 μL

sample

Waters BEH200 SEC protein standard

詳細

アナリシスノート

Arginine can have benefitial effects on peak shape and recovery of hydrophobic proteins when chromatographed by size exclusion chromatography. In this case, with protein standards that are not particularly hydrophobic, no dramatic affect is noted, as assessed by peak efficiency, peak widths, or peak retention.

法的情報

Zenix is a registered trademark of Sepax Technologies