コンテンツへスキップ
Merck

HPLC Analysis of mAb Aggregates on TSKgel® UltraSW Aggregate: Effect of Mobile Phase Additives on Resolution

HPLC Analysis of mAb Aggregates on TSKgel® UltraSW Aggregate: Effect of Mobile Phase Additives on Resolution application for HPLC

材料

ガードカラム

製品番号
詳細
価格

TSKgel® UltraSW Aggregate Guard Column

phase diol, L × I.D. 4 cm × 6 mm, 3 μm particle size

分析カラム

製品番号
詳細
価格

TSK-GEL®サイズ排除(SWタイプ)HPLCカラム

phase diol, L × I.D. 30 cm × 7.8 mm, 3 μm particle size

CONDITIONS

column

TSKgel UltraSW Aggregate 30 cm x 7.8 mm I.D., 3 μm particles (822856)

mobile phase

0.1 mol/L sodium phosphate buffer, pH 6.7, with an addition of either 0.2 mol/L arginine (A), or 0.2 mol/L proline (B)

flow rate

1.0 mL/min

column temp.

ambient

detector

UV, 280 nm

injection

20 μL

sample

mAB, aggregated at 75 °C for 5 min

詳細

アナリシスノート

While particle size, packing quality and column length are all important considerations in optimizing mAb aggregate analysis, mobile phase additives can also contribute to improving your separation. The addition of amino acids such as lysine and arginine to the mobile phase, increase aggregate recovery, while glutamine, proline and arginine improve monomer aggregate resolution, as well as monomer fragment resolution. Although the increases in resolution are not drastic, they confirm that increased resolution due to the use of an advanced mobile phase is possible and that mobile phase testing can contribute to a more reliable and robust aggregate analysis.

法的情報

TSKgel is a registered trademark of Tosoh Corporation