コンテンツへスキップ
Merck

GC Analysis of Volatiles in Purple Nitrile Gloves on SPB®-HAP after Collection/Desorption using Carbotrap® 300 Tube

GC Analysis of Volatiles in Purple Nitrile Gloves on SPB®-HAP after Collection/Desorption using Carbotrap® 300 Tube suitable for GC, application for air monitoring

材料

関連製品

製品番号
詳細
価格

ATIS吸着チューブインジェクターシステム 110 V

ATIS Adsorbent Tube Injector System 230V

ATIS抽出用ガラス器具(マイクロコネクター付き)

CONDITIONS

sample/matrix

0.5 g purple nitrile glove

thermal desorption tube

Carbotrap® 300 Glass-Fritted Thermal Desorption Tube, 1/4 in O.D. x 3.5 in L (30141-U)

desorption process

350 °C with 10 min hold

transfer line/valve temp.

180 °C

column

SPB®-HAP, 60 m x 0.32 mm I.D., 4 μm (25020-U)

oven

40 °C (3 min), 12 °C/min to 200 °C (20 min)

detector

MSD

MSD interface

210 °C

carrier gas

helium

dry purge

15 min