コンテンツへスキップ
Merck

G005753

HPLC Analysis of an Energy Drink on Ascentis® Express HILIC (UV Detection)

application for HPLC

ログイン組織・契約価格を表示する


About This Item

テクニック

HPLC: suitable

試験パラメータ

column: Ascentis Express HILlC, 10 cm x 3.0 mm I.D., 2.7 μm particles (53970-U)
mobile phase: [A] 100 mM ammonium acetate, pH 5.0 with acetic acid; [B] water; [C] acetonitrile; (09:01:90, A:B:C)
flow rate: 0.6 mL/min
pressure: 812 psi (56 bar)
column temp.: 35 °C
detector: UV, 254 nm
injection: 2 μL
sample: Dilute 1:9 in acetonitrile

適合性

application for HPLC

アプリケーション

food and beverages

アナリシスノート

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of common drink additives used in caffeinated energy drinks. The analysis was performed with UVand ELSD detectors in series, this chromatogram being from the UV.

法的情報

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

アナライト

詳細

    Caffeine
    Benzoic acid
    Sorbic acid

Choose from one of the most recent versions:

試験成績書(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact カスタマーサポート

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.

製品に関するお問い合わせはこちら(テクニカルサービス)