コンテンツへスキップ
Merck

KGW1A1TTT1

Millipore

Polysep II 1.0/0.2/0.1 µm(公称)、オプティキャップ(Opticap)® XLTカプセル

Opticap® XLT 10, pore size 1.0/0.2/0.1 μm, cartridge nominal length 10 in. (25 cm)

ログイン組織・契約価格を表示する


About This Item

UNSPSCコード:
23151806

物質

borosilicate glass fiber (BGF) membrane
mixed cellulose esters (MCE) membrane
polypropylene filter
polypropylene housing
polypropylene support
silicone seal

品質水準

法規制遵守

FDA 21CFR177-182 (all component materials)

無菌性

non-sterile

滅菌適合性

autoclavable compatible

製品種目

Opticap® XLT 10

特徴

hydrophilic

メーカー/製品名

Opticap®

パラメーター

1 bar max. inlet pressure (15 psi) at 80 °C
2.75 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
3.5 bar max. differential pressure (50 psid) at 25 °C
5.5 bar max. inlet pressure (80 psi) at 23 °C
80 psig max. inlet pressure

テクニック

prefiltration: suitable

カートリッジの長さ(公称)

10 in. (25 cm)

device L

37.6 cm (14.8 in.)

device size

10 in.

filter filtration area

0.42 m2

入口接続部直径

1-1/2 in.

inlet to outlet width

15.2 cm (6.0 in.)

出口接続部直径

1-1/2 in.

不純物

≤0.5 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.5 EU/mL bacterial endotoxins (LAL test, sample aqueous extraction)

マトリックス

Polysep II

pore size

1.0/0.2/0.1 μm nominal pore size
1.0/0.2/0.1 μm pore size

フィッティング

(38 mm (1 1/2 in.) Sanitary Flange Inlet and Outlet)

類似した製品をお探しですか? 訪問 製品比較ガイド

詳細

Device Configuration: Capsule

特徴および利点

Triple Layer

包装

Double Easy-Open bag

調製ノート

Sterilization Method:
3 autoclave cycles of Sterilization Method: 30min at 121°C; not in-line steam sterilizable

This product was manufactured with a media combination which meets the criteria for a "non-fiber releasing" filter as defined in 21CFR210.3(b)(6).

アナリシスノート

Gravimetric Extractables: ≤ 145 mg per capsule after a 5 L flush and 24 hours in water at controlled room temperature.
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥ 5 L

その他情報

Directions for Use:

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer first page of Visual Inspection Guide and introduction section of Mixed cellulose esters Tech Brief
  • Storage Statement: Please refer Mixed cellulose esters Tech Brief
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

法的情報

OPTICAP is a registered trademark of Merck KGaA, Darmstadt, Germany
Polysep is a trademark of Sigma-Aldrich Co. LLC

適切な製品が見つかりませんか。  

製品選択ツール.をお試しください


試験成績書(COA)

製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.

製品に関するお問い合わせはこちら(テクニカルサービス)