おすすめの製品
詳細
The Sterikon® plus Bioindicator contains a nutrient broth, sugar, a pH indicator, and spores of a non-pathogenic organism, Geobacillus stearothermophilus ATCC 7953 (sporulation optimized) in a single ampoule. The spores are completely killed upon heating in compressed steam at a temperature of 121 ± 0.5°C (245 kPa), after 15 minutes. At lower temperatures or lower exposure times, a small number of spores can survive at and can grow upon incubation. Sterikon® ampoules are placed into the autoclave along with the batch to be autoclaved. The efficiency of the sterilization process is checked by the incubation of the ampoules. Red (or red-violet) color of the ampoule contents indicates no growth of Geobacillus stearothermophilus and adequate sterilization, whereas the yellow (or yellow-orange) color of the ampoule contents indicates growth and inadequate sterilization.
アプリケーション
Sterikon® plus Bioindicator System is used to monitor the effectiveness of the autoclaving by steam sterilization for 15 minutes at 121°C. It is suitable for checking the sterilization efficiency of the culture media products used for testing of pharmaceuticals (especially drugs in ampoule form), food and water samples, along with canned foods and other products to be autoclaved.
関連事項
Replaces: 1.10274.0003
アナリシスノート
Appearance: opalescent and reddish violet
Number of viable spores per ampoule: 5.0E+05 - 1.0E+07
Growth (35 °C): no growth
Growth (60 °C): growth
Performance test (6 min, 121 °C): Survival time 6 min/121 °C
Performance test (15 min, 121 °C): Kill time 15 min/121 °C
D value (121 °C): 1.5 - 2.0 min
Z value: 7.0 - 10.0 °C
Identity: Geobacillus stearo-thermophilus ATCC 7953
The determination of the D-value is based on the Stumbo-Murphy-Cochran method.
The identity control is carried out via MicroSeq-System.
Organism: Geobacillus stearothermophilus ATCC 7953 (sporulation optimized).
Ampoules: 2 ml.
Storage temperature: +2° to +8°C.
Number of viable spores per ampoule: 5.0E+05 - 1.0E+07
Growth (35 °C): no growth
Growth (60 °C): growth
Performance test (6 min, 121 °C): Survival time 6 min/121 °C
Performance test (15 min, 121 °C): Kill time 15 min/121 °C
D value (121 °C): 1.5 - 2.0 min
Z value: 7.0 - 10.0 °C
Identity: Geobacillus stearo-thermophilus ATCC 7953
The determination of the D-value is based on the Stumbo-Murphy-Cochran method.
The identity control is carried out via MicroSeq-System.
Organism: Geobacillus stearothermophilus ATCC 7953 (sporulation optimized).
Ampoules: 2 ml.
Storage temperature: +2° to +8°C.
法的情報
STERIKON is a registered trademark of Merck KGaA, Darmstadt, Germany
保管分類コード
12 - Non Combustible Liquids
WGK
WGK 1
引火点(°F)
Not applicable
引火点(℃)
Not applicable
適用法令
試験研究用途を考慮した関連法令を主に挙げております。化学物質以外については、一部の情報のみ提供しています。 製品を安全かつ合法的に使用することは、使用者の義務です。最新情報により修正される場合があります。WEBの反映には時間を要することがあるため、適宜SDSをご参照ください。
消防法
第4類:引火性液体
アルコール類
危険等級II
試験成績書(COA)
製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。