G005922
HPLC Analysis of Aflatoxins B1, B2, G1 and G2 in Corn on Ascentis® Express C18 after SPE using Supel™ Tox AflaZea and Fluorescence Detection
application for HPLC
About This Item
tecniche
HPLC: suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 25 g corn meal in 100 mL acetonitrile:water (84:16)
SPE tube/cartridge: Supel™ Tox AflaZea, 6 mL (55314-U)
sample addition: 2 mL
elution: collect sample in 5 mL silanized glass vial
eluate post-treatment: transfer 200 μL sample, add 300 μL TFA derivatization solution (70:20:10 water:TFA:acetic acid), heat in microreactor 20 min at 65 °C, cool, add 580 μL deionized water, vortex
column: Ascentis Express C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (53825-U)
mobile phase: [A] water; [B] acetonitrile; (80:20, A:B)
flow rate: 0.350 mL/min
pressure: 325 psi (22.4 bar)
column temp.: 35 °C
detector: fluorescence, ex 360 nm, em 440 nm
injection: 20 μL
Compatibilità
application for HPLC
applicazioni
food and beverages
Risultati analitici
Note legali
Analita
Colonna analitica
Prodotti correlati
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documenti section.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.