G003766
HPLC Analysis of Diazepam and Metabolites in Serum on Discovery® C18 after SPE using Discovery® DSC-8
application for SPE, application for HPLC
About This Item
tecniche
HPLC: suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: goat serum spiked with diazepam and metabolites at 0.5 μg/mL and diluted 1:1 in 25 mM ammonium formate, pH 7.1
SPE well plate: Discovery DSC-8 96-well SPE Plate, 100 mg/well (575627-U)
condition: 1 mL methanol followed by 1 mL DI water.
sample addition: 1 mL spiked goat serum
washing: 1 mL 25% methanol in 25 mM ammonium formate, pH 2.75
elution: 1 mL 60% methanol in 25 mM ammonium formate, pH 2.75
elution: Evaporate eluate with nitrogen purge (30 °C; ~15 min.). Reconstitute in 200 μL mobile phase.
column: Discovery C18, 5 cm x 4.6 mm I.D., 5 μm particles (504947)
mobile phase: [A] Methanol; [B]10 mM ammonium acetate, pH 4.5;(45:55, A:B)
flow rate: 1.5 mL/min
column temp.: 35 °C
detector: UV, 240 nm
injection: 25 μL
Compatibilità
application for HPLC
application for SPE
applicazioni
clinical
forensics and toxicology
Note legali
Colonna analitica
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documenti section.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.