Stericup® Quick Release, or “QR”, filters feature a connection between funnel filter and receiver bottle that requires only a quarter-turn release to facilitate funnel removal. These filters come fully engaged – there is no need to tighten before use. The receiver bottle cap has flanges on the top surface to improve grip, as well as a positive lock stop to ensure a leak-proof seal with the receiver bottle. Other improvements include a large, frosted writing surface on the receiver bottle and a lighter cap color to facilitate labelling, more prominent volume markings on the filtration funnel and receiver bottle, and a product label with a peel-off lot record strip on the bag for easy viewing in storage.
Scegli un formato
| A voi/SKU | Disponibilità | Prezzo |
|---|---|---|
12 units | Disponibile per la spedizione OGGIdaMerck Life Science S.r.l. | 145,00 € |
Informazioni su questo articolo
Quality Segment
material
polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing
product line
Stericup®
feature
hydrophilic
packaging
pack of 12 ea
parameter
45 °C max. temp.
description
2.4 mL (hold-up volume)
technique(s)
sterile filtration: suitable
H
120 mm
diam.
12.5 cm
filter capacity
150 mL
filter diam.
73 mm
filtration area
40 cm2
matrix
Express PLUS™
pore size
0.22 μm pore size
fitting
inlet funnel, outlet flask (Receiver)
shipped in
ambient
storage temp.
no temp limit
General description
Si servono di affidabili membrane dalle alte prestazioni per un'ottimale filtrazione sterilizzante di reagenti, tamponi e terreni di laboratorio. Questi dispositivi uniscono portate e rendimenti eccellenti a un ridotto adsorbimento aspecifico e a un design stabile e smussato.
Questo sistema Stericup® per il trattamento e la conservazione di soluzioni, combina un dispositivo filtrante da bottiglia di 150 mL di capacità con membrana in polietersolfone (PES) da 0,22 μm, un flacone di raccolta da 150 mL e un tappo.
Application
- Terreni di coltura +/- additivi
- Tamponi
- Soluzioni biologiche
Features and Benefits
- L'imbuto filtrante “Quick Release” si sgancia dal flacone di raccolta semplicemente con una rotazione di un quarto di giro, riducendo il rischio di fuoriuscite e semplificando la manipolazione sul banco e sotto cappa.
- L'identificazione del prodotto, ben visibile sull'imbuto filtrante Steritop®, espone chiaramente le caratteristiche della membrana quali composizione e dimensione dei pori, consentendo di scegliere rapidamente l'unità appropriata per la propria applicazione.
- Il tappo "click seal" è flangiato e modificato per consentire una presa sicura anche se bagnato; inoltre, un segnale di stop tattile conferma la chiusura del flacone di raccolta. Diminuiranno, pertanto, le preoccupazioni circa la possibilità di rovesciare o di contaminare il contenuto sterilizzato.
- L'ampia superficie di scrittura satinata e il tappo di colore più chiaro del flacone di raccolta facilitano una chiara etichettatura, migliorando la leggibilità e riducendo il rischio di sbavature e la conseguente impossibilità di lettura.
- Il tappo "click seal" può essere appoggiato sul fianco, evitando rischi di contaminazione.
- I flaconi di raccolta sono impilabili e consentono, quindi, di risparmiare spazio di stoccaggio.
Legal Information
1 of 1
Questo articolo | |||
|---|---|---|---|
| description pore size 0.22 μm, polyethersulfone membrane, filter capacity 150 mL, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 1000 mL, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 500 mL, pack of 12 ea | description pore size 0.22 μm, polyethersulfone membrane, filter capacity 250 mL, pack of 12 ea |
| pore size 0.22 μm pore size | pore size 0.22 μm pore size | pore size 0.22 μm pore size | pore size 0.22 μm pore size |
| filter capacity 150 mL | filter capacity 1000 mL | filter capacity 500 mL | filter capacity 250 mL |
| material polyethersulfone membrane, polystyrene funnel(s), polystyrene bottle, polystyrene housing | material polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing | material polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing | material polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing |
| fitting inlet funnel, outlet flask (Receiver) | fitting inlet funnel, outlet flask (Receiver) | fitting inlet funnel, outlet flask (Receiver) | fitting inlet funnel, outlet flask (Receiver) |
| feature hydrophilic | feature hydrophilic | feature hydrophilic | feature hydrophilic |
| packaging pack of 12 ea | packaging pack of 12 ea | packaging pack of 12 ea | packaging pack of 12 ea |
Still not finding the right product?
Explore all of our products under Sistema di filtrazione sotto vuoto Stericup® Quick Release Millipore®
Scegli una delle versioni più recenti:
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
-
What is the difference between classic Stericup® filters and Stericup® “Quick Release” filters?
1 risposta-
Utile?
-
-
Can Stericup® and Steritop® filters be reused?
1 risposta-
All Stericup® and Steritop® filters were developed for single use, disposable processing.
Utile?
-
-
Can we filter oil-type liquids with Stericup® filters?
Can we filter oil-type liquids with Stericup® filters?
1 risposta-
Likely not, due to pore size/possible fouling/chemical compatibility. Refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
Utile?
-
-
What is the maximum vacuum pressure for filtration?
1 risposta-
700 mm Hg (27.5” Hg, 0.9 bar) differential vacuum at 25 °C (77 °F)
Utile?
-
-
What is the maximum processing volume for Stericup® and Steritop® filters? Can I add more volume after initial filtration?
1 risposta-
All Stericup® and Steritop® filters were developed for single-use application. The maximum volume is dependent on the receiver bottle capacity. Stericup® filters come with a receiver bottle with a specified maximum volume. For Steritop® filters, the maximum recommended volume is 2 L with an appropriate-sized receiver bottle. Larger processing volumes can result in membrane fouling, increase chance of spillage and contamination, and lead to possible airlock*.
*Note that devices with PES membranes can airlock if filtered to dryness. Some liquid must be left on top of the filter, then the system stopped, receiver changed or emptied, and more sample volume added before restart. This increases the chance of spillage.Utile?
-
-
Are Stericup® and Steritop® filters GMP-certified?
1 risposta-
No, Stericup® and Steritop® devices are not GMP-approved.
Utile?
-
-
What tubing can be used with Stericup® and Steritop® vacuum filters?
What tubing can be used with Stericup® and Steritop® vacuum filters?
1 risposta-
Stericup® and Steritop® filters are fitted with a 1/4” tube adapter. We recommend 3/16 in. (4.8 mm) ID silicone tubing to fit this tube adapter: https://www.sigmaaldrich.com/product/mm/xx7100004
Utile?
-
-
What leachables are detected during filtration processes using Stericup® filters?
1 risposta-
We have not performed extractables/leachables studies for our Stericup® filters.
Utile?
-
-
The User Guide says “Use only glass or plastic bottles designed for vacuum applications”. Can you specify which bottles are compatible?
1 risposta-
Steritop® Quick Release bottle top units require a vacuum-rated receiver bottle with either a GL33 or GL45 mm closure thread as specified. Steritop® E bottle top units require a vacuum-rated receiver bottle with a GL45 mm closure thread. Bottles no larger than 2 L should be used.
Glass bottles, GL33:
• Wheaton media bottles (glass), 500 mL (Product Z364886)
• Wheaton media bottles (glass), 1 L (Product Z364894)
Glass bottles, GL45
• Corning® glass media storage bottle, 125 mL (Product CLS1399125)
• Corning® glass media storage bottle, 250 mL (Product CLS1399250)
• Corning® glass media storage bottle, 500 mL (Product CLS1399500)
• Corning® glass media storage bottle, 1 L (Product CLS13991L)
• Pyrex® round media storage bottle, 500 mL (Product CLS1395500)
• Pyrex® round media storage bottle, 1 L (Product CLS13951L)
• KIMBLE® KIMAX® Media bottle, 250 mL (Product DWK14395250)
• KIMBLE® KIMAX® Media bottle, 500 mL (Product DWK14395500)
• KIMBLE® KIMAX® Media bottle, 1 L (Product DWK143951000)
Plastic (pre-sterile) bottles, GL45:
• Stericup® Quick Release Receiver flask, 250 mL (Product S200B02RE)
• Stericup® Quick Release Receiver flask, 500 mL (Product S200B05RE)
• Stericup® Quick Release Receiver flask, 1000 mL (Product S200B10RE)Utile?
-
-
Can Stericup® and Steritop® filters be used to prepare solutions that will be administered to patients?
1 risposta-
No, all Stericup® and Steritop® devices are Research Use Only.
Utile?
-