跳转至内容
Merck

UHPLC/MS Analysis of Diazepam and Metabolites on Titan C18, Comparison of Flow Rate and Column Length

UHPLC/MS Analysis of Diazepam and Metabolites on Titan C18, Comparison of Flow Rate and Column Length application for UHPLC

材料

標準

产品编号
说明
价格

奥沙西泮标准液 溶液

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

去甲地西泮标准液 溶液

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

地西泮标准液 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

替马西泮标准液 溶液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Titan C18, 2 cm x 2.1 mm I.D., 1.9 μm particles (577120-U)

mobile phase

[A] 0.1% formic acid; [B] acetonitrile; (60:40, A:B)

flow rate

various

temp.

35 °C

detector

TIC MRM (UV also runs in series)

injection

1 μL

sample

100 ng/mL in mobile phase

说明

分析報告

These two sets of chromatograms show separations on 100, 50, 30, and 20 mm x 2.1 mm I.D. columns and at different flow rates and demonstrate the utility of the 1.9 μm Titan particles. The highest grade solvents provided clean, robust operation. Cerilliant CRMs provided reliable quantification.

法律資訊

null