跳转至内容
Merck

LC/MS/MS Analysis of Aflatoxins in Hops on Ascentis® Express 2.7 μm Phenyl-Hexyl after Cleanup Using Supel Tox AflaZea SPE

LC/MS/MS Analysis of Aflatoxins in Hops on Ascentis® Express 2.7 μm Phenyl-Hexyl after Cleanup Using Supel™ Tox AflaZea SPE application for LC-MS, application for SPE

材料

分析柱

产品编号
说明
价格

Ascentis® Express Phenyl-Hexyl,2.7μm HPLC色谱柱

2.7 μm particle size, L × I.D. 5 cm × 2.1 mm

標準

产品编号
说明
价格

黄曲霉毒素混合物 4 溶液

analytical standard

附屬品

产品编号
说明
价格

黑色塑料盖

Solid top, white rubber liner, thread for 8-425, pkg of 1000 ea

Visiprep SPE 真空固相萃取装置

DL (Disposable Liner), 12-port model

Visiprep SPE 真空固相萃取装置

DL (Disposable Liner), 24-port model

经硅烷化处理螺纹口样品瓶

volume 2 mL, amber glass vial, O.D. × H × I.D. 12 mm × 32 mm × 4.6 mm

QuEChERS Shaker and Rack Starter Kit

AC/DC input 115 V AC (USA Compatible Plug)

CONDITIONS

sample/matrix

hops extract was spiked with aflatoxins at corresponding concentrations of 6 ppb each (for Aflatoxin B2 and G2) and 24 ppb each (for Afltoxin B1 and G1) in hops

SPE tube/cartridge

Supel Tox AflaZea SPE, 6 mL (55314-U)

extraction process

0.5 g of finely ground hops is mixed with 10 mL of 84:16 acetonitrile:water for 30 minutes, centrifuged

sample addition

2 mL of hops sample extract

elution

apply strong vacuum and collect the sample into plastic tube

sample preparation

dilute 0.2 mL of the extract with 0.8 mL distilled water and mix well

column

Ascentis Express Phenyl-Hexyl, 5 cm x 2.1 mm I.D., 2.7 μm particles (53334-U)

mobile phase

[A] 5 mM ammonium formate with 1% formic acid in water; [B] 5 mM ammonium formate with 1% formic acid in methanol

gradient

30% B to 60% B in 3 min; to 100% B in 2 in; held at 100% B for 2 min; to 30% B in 0.1 min; held at 30% B for 1.4 min

flow rate

0.4 mL/min

pressure

5511 psi (380 bar)

column temp.

40 °C

detector

MS, ESI(+), MRM 331.3/189.0, 329.1/243.0, 315.9/259.0, 313.1/241.0

injection

10 μL

说明

分析報告

The complex composition of hops makes it difficult to apply sample cleanup methods commonly used for other commodities. In this application, interference removal on Supel Tox AflaZea SPE was used for sample cleanup providing quick and effective sample preparation prior to separation on an Ascentis Express Phenyl-Hexyl column and mass spectrometric detection. It is recommended to use silanized glass vials for this analysis.

其他說明

Use silanized vials for analysis.

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC