跳转至内容
Merck

G005644

HPLC Analysis of Nucleosides on Ascentis® Express OH5 (HILIC Mode)

application for HPLC

登录查看公司和协议定价


About This Item

技術

HPLC: suitable

測試參數

column: Ascentis Express OH5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53757-U)
mobile phase: [A] 5 mM ammonium acetate, pH 5.0 with acetic acid in 95:5, acetonitrile:water; [B] 5 mM ammonium acetate, pH 5.0 with acetic acid in 80:20, acetonitrile:water
gradient: 0% B held for 1 min; to 100% B in 10 min; held at 100% B for 1 min
flow rate: 0.3 mL/min
column temp.: 25 °C
detector: UV, 250 nm
injection: 2 μL
sample: 10 - 100 μg/mL in 95:5, acetonitrile:water (pH of buffer stock (in water) was adjusted before further dilution with water and/or acentonitrile)

適合性

application for HPLC

應用

life science and biopharma

分析報告

Nucleosides are involved in many aspects of metabolism. Analogs of nucleosides have medical application in antivirals or anticancer treatments. This application demonstrates the utility of hydrophilic interaction chromatography (HILIC) for the retention and resolution of nucleosides and analogs thereof. Elution is not solely explained by logD but by other ionic and solvophobic effects.

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    5-Methyluridine
    DMT-2′O-TC-rU Phosphoramidite
    2-Thiocytidine dihydrate
    2′-O-Methylcytidine
    Inosine
    5-Methylcytidine
    Cytidine
    DMT-2′O-TC-rG(ib) Phosphoramidite
    1-Methyladenosine
    7-Methylguanosine
    查看所有结果 (10)

Choose from one of the most recent versions:

分析证书(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact 客户支持

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门