G005854
LC/MS Analysis of Antiarrhythmic Drugs and Metabolites in Plasma on Ascentis® Express HILIC following Sample Prep using HybridSPE®-Phospholipid
application for HPLC
About This Item
technique(s)
HPLC: suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rabbit plasma, unfiltered K2-EDTA, spiked with each compound at 100 ng/mL (3:1, plasma:1% formic acid in acetonitrile)
SPE tube/cartridge: HybridSPE-Phospholipid, 96-well plate (575656-U)
sample addition: to each well add 100 μL plasma, followed by a 300 μL of 1% formic acid in acetonitrile
elution: collect, filtrate and analyze directly.
column: Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)
mobile phase: [A] 5 mM ammonium formate; [B] 5 mM ammonium formate in acetonitrile, 5:95 (A:B), adjusted to pH 7.0 with formic acid
flow rate: 0.4 mL/min
pressure: 1305 psi (90 bar)
column temp.: 35 °C
detector: ESI+, 100-1000 m/z
injection: 0.5 μL
suitability
application for HPLC
application(s)
clinical
Analysis Note
Legal Information
Analyt
SPE tube or plate
analytical column
standard
Vyberte jednu z posledních verzí:
Osvědčení o analýze (COA)
Je nám líto, ale pro tento produkt momentálně nemáme COA k dispozici online.
Potřebujete-li pomoc, obraťte se na Zákaznická podpora
Již tento produkt vlastníte?
Dokumenty související s produkty, které jste v minulosti zakoupili, byly za účelem usnadnění shromážděny ve vaší Knihovně dokumentů.
Náš tým vědeckých pracovníků má zkušenosti ve všech oblastech výzkumu, včetně přírodních věd, materiálových věd, chemické syntézy, chromatografie, analytiky a mnoha dalších..
Obraťte se na technický servis.