Saltar al contenido
Merck

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

Materiales

Fibra para SPME

Referencia del producto
Descripción
Precios

SPME fiber assembly polyacrylate (PA)

df 85 μm(PA, for use with manual holder, needle size 24 ga

columna analítica

Referencia del producto
Descripción
Precios

Equity-1

L × I.D. 30 m × 0.32 mm, df 1.00 μm

CONDITIONS

sample/matrix

2 g of B100 biodiesel containing 0.01% methanol and 0.0785% 2-propanol in a 4 mL vial

SPME fiber

PK3 SPME ASSY 85UM POLYACRYLATE FS 24GA MAN (WHITE) (57304)

extraction

headspace, 60 °C for 5 min

desorption process

250 °C for 0.75 min

column

Equity-1, 30 m x 0.32 mm I.D., 1.0 μm (28057-U)

oven

50 °C

inj. temp.

250 °C

detector

FID, 200 °C

carrier gas

helium, 30 cm/sec

liner

0.75 mm I.D., SPME type, straight design (unpacked)

Descripción

Información legal

null