In Vitro Toxicology Assay Kit, Lactic Dehydrogenase based

Sinónimos:

LDH cytotoxicity assay

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

Código UNSPSC:
12352207
NACRES:
NA.84
En este momento no podemos mostrarle ni los precios ni la disponibilidad

Formulario

liquid

Nivel de calidad

uso

 kit sufficient for 500 tests

envase

pkg of 1 kit

condiciones de almacenamiento

dry at room temperature

técnicas

protein quantification: suitable

λmáx.

490 nm

aplicaciones

cell analysis
detection

método de detección

colorimetric

temp. de almacenamiento

−20°C

Descripción general

Tox7 is a member of Sigma′s CytoToxicity Kit series. It is based on the measurement of lactate dehydrogenase (LDH) release from cells as a measure of cell membrane damage and cytotoxicity. This technique has been utilized as an alternative to 51Cr release for cell mediated cytotoxicity assays, as well as conventional cytotoxicity resulting from interaction of a test material with the cell.

Aplicación

In Vitro Toxicology Assay Kit, Lactic Dehydrogenase based has been used:
  • to detect lactate dehydrogenase release in human umbilical vein endothelial cells (HUVECs) and human cardiac myocytes (HCMs) for lactate dehydrogenase (LDH) assay[1]
  • to determine the activity of lactate dehydrogenase leakage of MDA-MB-231 cells[2]
  • to examine the activity of LDH in RAW 264.7 cells [3], in LDH cytotoxicity assay in human cell lines[4]
  • for spectrophotometric measurement of viable cells. Absorbance of converted dye is measured at a wavelength of 490 nm.

Acciones bioquímicas o fisiológicas

The lactate dehydrogenase assay is a means of measuring either the number of cells via total cytoplasmic lactate dehydrogenase (LDH) or membrane integrity as a function of the amount of cytoplasmic LDH released into the medium. This method is simple, accurate and yields reproducible results. The assay is based on the reduction of NAD by LDH. The resulting reduced NAD (NADH) is utilized in the stoichiometric conversion of a tetrazolium dye to a soluble, colored formazan derivative.
.

Frases de peligro

Consejos de prudencia

Clasificaciones de peligro

Aquatic Chronic 3

Código de clase de almacenamiento

10 - Combustible liquids


Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

¿No ve la versión correcta?

Si necesita una versión concreta, puede buscar un certificado específico por el número de lote.

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Katarzyna Walczak et al.
International journal of molecular sciences, 21(21) (2020-10-30)
Tryptophan metabolites: kynurenine (KYN), kynurenic acid (KYNA) and 6-formylindolo[3,2-b]carbazole (FICZ) are considered aryl hydrocarbon receptor (AhR) ligands. AhR is mainly expressed in barrier tissues, including skin, and is involved in various physiological and pathological processes in skin. We studied the
Effects of 5-fluorouracil on morphology, cell cycle, proliferation, apoptosis, autophagy and ROS production in endothelial cells and cardiomyocytes
Focaccetti C, et al.
Testing, 10, e0115686-e0115686 (2015)
Christian Polley et al.
Materials (Basel, Switzerland), 13(7) (2020-04-15)
The prevalence of large bone defects is still a major problem in surgical clinics. It is, thus, not a surprise that bone-related research, especially in the field of bone tissue engineering, is a major issue in medical research. Researchers worldwide
Cytotoxicity of biologically synthesized silver nanoparticles in MDA-MB-231 human breast cancer cells
Gurunathan S, et al.
BioMed Research International (2013)
Riboflavin deprivation inhibits macrophage viability and activity-a study on the RAW 264.7 cell line
Mazur-BAI, et al.
The British Journal of Nutrition, 110(3), 509-514 (2013)

Artículos

Quality control guidelines to maintain high quality authenticated and contamination-free cell cultures. Free ECACC handbook download.

Quality control guidelines to maintain high quality authenticated and contamination-free cell cultures. Free ECACC handbook download.

Quality control guidelines to maintain high quality authenticated and contamination-free cell cultures. Free ECACC handbook download.

Quality control guidelines to maintain high quality authenticated and contamination-free cell cultures. Free ECACC handbook download.

Ver todo

Questions

1–10 of 10 Questions  
  1. in the TOX7, it describes that the 25 ml of 1× LDH Assay Cofactor Preparation by adding 25 ml of tissue culture grade water to the bottle of lyophilized cofactors (Catalog Number L2527). Can I use Sterile Filtered and Autoclaved Sterile water to prepare?

    1 answer
    1. The recommendation to use tissue culture grade water is primarily to ensure there are no trace elements or contaminants present that could impact the metabolic health of the cells being screened. The incubation period is relatively brief and cells are not likely to be negatively impacted. The use of sterile filtered or autoclaved water will not negatively impact the functionality of the kit. It would be up to the end user to deviate from the recommended protocol.

      Helpful?

  2. How do i estimate the LDH release? what is the equation?

    1 answer
    1. The TOX7 kit is utilized primarily as a cell viability assay based upon cellular metabolic activity. LDH reduces NAD+ which is then converted to a tetrazolium dye. This dye is then solubilized and read at 490nm. The resulting absorbance in associated with cell density or viability.

      Please refer to the Product Information Sheet for more information regarding this product: https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/336/418/tox7bul.pdf

      Helpful?

  3. Can TOX7, In Vitro Toxicology Assay kit be used to quantitate the amount of LDH in my samples?

    1 answer
    1. This assay kit was designed as a semi-quantitative assay for LDH. The change in the OD between the treated samples and control samples is to be compared and is generally reported as a fold increase/decrease from the control. The kit does not contain a LDH standard. Product L3916 can be used as an LDH standard for this assay. The detection range (LDH units/ml) for the kit has not been established.

      Helpful?

  4. The TOX7 dye solution has a precipitate present. Can I still use it?

    1 answer
    1. Product No. L2277 LDH Assay Dye Solution is a saturated solution and some precipitation may appear upon thawing. The product will partially re-solubilize when warmed to 37°C and the resulting solution can be decanted away from any remaining precipitate. Precipitated dye added to reactions, or removed from the solution will not affect the performance or results obtained from this kit.

      Helpful?

  5. Can you use tissue explants in product TOX7, In Vitro Toxicology Assay Kit, Lactic Dehydrogenase based?

    1 answer
    1. Unfortunately, we have not tested product TOX7, In Vitro Toxicology Assay Kit, on tissue samples.  If the tissue is minced well and can be lysed completely with the lysis buffer for the cells, it may work to determine cell biomass.  The following paper modified the assay for use with explants.Benyo, D.F., et al., "Hypoxia Stimulates Cytokine Production by Villous Explants from the Human Placenta", The Journal of Clinical Endocrinology & Metabolism 82(5), 1582-1588 (1997).Assessment of explant viability was routinely monitored by measuring the release of LDH into medium relative to a 1% Triton X-100 (Sigma)-lysed positive control. The assay employed a modification of the TOX7 viability kit supplied by Sigma-Aldrich and entailed adding a 20-fold volume excess of assay mixture (160 mL) to the medium sample tested (8 mL). The assay mixture consisted of equal volumes of substrate (lactate), enzyme (diaphorase), and dye (nitro blue tetrazolium) solutions, and the reaction was allowed to proceed for 20 min in a microtiter plate before the addition of 0.1 vol 1N HCl. Color change was quantified by reading the absorbance at 490 nm, and the percent cell death calculated.

      Helpful?

  6. Will Triton X100  interfere with the TOX7 In Vitro Toxicology Kit assay?

    1 answer
    1. Triton X100 will not interfere with TOX7 assay.

      Helpful?

  7. What is the minimum number of cells to use in the LDH assay, Product TOX7, In Vitro Toxicology Assay Kit?

    1 answer
    1. Using Product TOX7, In Vitro Toxicology Assay Kit, the minimum cell number for best response is 2 x 104.

      Helpful?

  8. What is the Department of Transportation shipping information for this product?

    1 answer
    1. Transportation information can be found in Section 14 of the product's (M)SDS.To access the shipping information for this material, use the link on the product detail page for the product.

      Helpful?

  9. I am using a protease inhibitor cocktail in the cell culture media.  Will this interfere with the TOX7 In Vitro Toxicology assay?

    1 answer
    1. The protease inhibitor cocktails do not interfere with LDH activity. Media from cells cultured with protease inhibitors can be used in this assay.

      Helpful?

  10. I'm using the TOX7 In Vitro Toxicology Assay Kit. I do not have a 690 nm filter to use for a reference/background. What can I use?

    1 answer
    1. The reference wavelength needs to be greater than the absorbance spectra of the assay. The greater the difference between the assay wavelength and the reference wavelength, the better.  A reference wavelength of 590 nm or greater can be used with the substrate for this assay.  If a reference wavelength is not available, the assay will work.  The use of the reference wavelength increases the signal to noise ratio to give a better assay range.

      Helpful?