Passa al contenuto
Merck

LC/MS Analysis of L-Methionine Sulfoximine and Sulfone on Ascentis® Express OH5

LC/MS Analysis of L-Methionine Sulfoximine and Sulfone on Ascentis® Express OH5 application for HPLC

Materiali

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Ascentis® Express OH5, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 3 mm

Standard

N° Catalogo
Descrizione
Determinazione del prezzo

L-Methionine sulfone

L-metionina sulfoximina

suitable for cell culture

CONDITIONS

column

Ascentis Express OH5, 10 cm x 3.0 mm I.D., 2.7 μm particles (53769-U)

mobile phase

[A] acetonitrile; [B] water with 0.1% formic acid, pH 3.5 with ammonium formate; (75:25, A:B)

flow rate

0.4 mL/min

pressure

1350 ppi (93 bar)

column temp.

25 °C

detector

MS, ESI (+), TIC, m/z 150-300

injection

5 μL

Descrizione

Risultati analitici

L-methionine sulfoximine and L-buthionine sulfoximine are used in are used to prevent additional enzyme activity in Chinese hampster ovary (CHO) cell lines with supplemental Glutamine Synthase (GS). It is of interest to be able to monitor both purity of such molecules in order to control addition to cell media as well as for the assay of parent molecules during use. The analytes are highly polar and thus should be amenable to HILIC-LC-MS analysis. This first installment investigates the ability of Ascentis Express HILIC and Ascentis Express OH5 stationary phases for the separation of the methionine analog and related compounds as well as methionine and buthionine separations.

Note legali

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany