Passa al contenuto
Merck

HPLC Analysis of Water Soluble Vitamins on Ascentis® RP-Amide

HPLC Analysis of Water Soluble Vitamins on Ascentis® RP-Amide application for HPLC

Materiali

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Ascentis® RP-Amide HPLC Column

5 μm particle size, L × I.D. 15 cm × 4.6 mm

Standard

N° Catalogo
Descrizione
Determinazione del prezzo

Thiamine hydrochloride

reagent grade, ≥99% (HPLC)

Pyridoxine

≥98%

Acido ascorbico

United States Pharmacopeia (USP) Reference Standard

Acido nicotinico

United States Pharmacopeia (USP) Reference Standard

CONDITIONS

column

Ascentis RP-Amide, 15 cm x 4.6 mm I.D., 5 μm particles (565324-U)

mobile phase

25 mM potassium phosphate monobasic, pH 3.5

flow rate

1.0 mL/min

column temp.

30 °C

detector

UV, 230 nm

injection

10 μL

Descrizione

Note legali

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany