Passa al contenuto
Merck

HPLC Analysis of Protein Standards on Zenix® SEC-300 at Low Flow Rate

HPLC Analysis of Protein Standards on Zenix® SEC-300 at Low Flow Rate application for HPLC

Materiali

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 7.8 mm, 3 μm particle size, stainless steel

Componente della fase mobile

N° Catalogo
Descrizione
Determinazione del prezzo

Acido fosforico

≥85 wt. % in H2O, ≥99.999% trace metals basis

Standard

N° Catalogo
Descrizione
Determinazione del prezzo

Myoglobin from equine heart

≥90% (SDS-PAGE), essentially salt-free, lyophilized powder

Uracil

≥99.0%

Thyroglobulin from bovine thyroid

powder, ≥90% (agarose gel electrophoresis)

IgG from rat serum

reagent grade, ≥95% (SDS-PAGE), essentially salt-free, lyophilized powder

Sieroalbumina

lyophilized powder, crystallized, ≥98.0% (GE)

CONDITIONS

column

Zenix SEC-300, 30 cm x 7.8 mm I.D., 3 μm (Z777033)

mobile phase

0.15 M potassium phosphate monobasic, pH 7.0 with phosphoric acid

flow rate

0.1 mL/min

pressure

173 psi (12 bar)

column temp.

25 °C

detector

UV, 215 nm

injection

4 μL

sample

protein standards

Descrizione

Risultati analitici

Due to the low diffusion coefficents of proteins, optimal flow in size exclusion chromatography is often very low. However, the disadvantage is long run times. This example demonstrates both features.

Note legali

Zenix is a registered trademark of Sepax Technologies