Passa al contenuto
Merck

GC Analysis of Sulfur Compounds in Gasoline on SPB®-1 SULFUR

GC Analysis of Sulfur Compounds in Gasoline on SPB®-1 SULFUR suitable for GC

Materiali

Colonna analitica

N° Catalogo
Descrizione
Determinazione del prezzo

SPB®-1 SULFUR Capillary GC Column

L × I.D. 30 m × 0.32 mm, df 4.00 μm

CONDITIONS

column

SPB-1 SULFUR, 30 m x 0.32 mm I.D., 4.0 μm (24158)

oven

35 ºC, 5 ºC/min to 200 ºC

inj. temp.

275 °C

detector

FID, 300 ºC; FPD, 300 ºC

carrier gas

helium, 20 cm/sec

injection

1 μL 100:1 split

sample

Gasoline

Descrizione

Risultati analitici

Two issues must be overcome for the analysis of sulfur compounds in fuels. 1) Detection of these compounds in a hydrocarbon sample matrix, and 2) adsorption of these compounds by the column. Sulfur specific detectors, such as a flame photometric detector (FPD), can be used to overcome the first challenge. A low phase ratio column, such as the SPB-1 SULFUR, can be used to overcome the second challenge. This slide shows the analysis of a gasoline sample, in which the column effluent was split to two detectors. The top chromatogram shows the FID response, and the bottom chromatogram shows the FPD response.

Note legali

SPB is a registered trademark of Merck KGaA, Darmstadt, Germany